Azbil Corporation
Installation
MagneW
FLEX+/PLUS+
Electromagnetic Flowmeter Detector
2-33
2.5.4: Installation on a metal pipe using SUS grounding
ring material
The installation method described in this section is for installation on a metal pipe
using SUS grounding ring material. For the installation method corresponding to any
other combination of materials, refer to Table 2-8 on page 2-32.
For installation using SUS grounding ring material, Azbil Corporation recommends
non-rubber gaskets such as joint sheet or PTFE. See Table 2-7 on page 2-31 for
recommended inner (bore) diameters.
To install a flanged detector on a metal pipe using SUS grounding ring material
and a specified torque:
• Install the detector as shown in the following illustration. The torque level for
tightening the bolts is not related to the gasket material.
See Table 2-6 on page 2-26 for the appropriate torque.
See Table 2-7 on page 2-31 for the inner diameter of the gasket.
Lining
Grounding ring
Gasket
11 mm
(0.43 inch)
Figure 2-24 Installation on a metal pipe using SUS grounding ring material and
a specified torque