![Azbil MagneW FLEX+ User Manual Download Page 21](http://html1.mh-extra.com/html/azbil/magnew-flex/magnew-flex_user-manual_3042665021.webp)
Azbil Corporation
Introduction
MagneW
FLEX+/PLUS+
Electromagnetic Flowmeter Detector
1-7
1.3.3: Union / Hose / Clamp type detector
Grounding terminal
Waterproof
gland
Flow direction
mark
Electrode
cover
Hose
Clamp
Union
Mounting screw(4 places)
Terminal box
Terminal box cover
Figure 1-6 Union / Hose / Clamp detector main parts
Name
Function
Flow direction mark
Indicates the direction of fluid flow.
Mount the detector so that the measured fluid flows in the
direction indicated by this mark.
Electrodes
The electrodes generate an electromotive force signal
proportional to the flow rate of the fluid passing through
the detector. The electrode material varies depending on
the corrosion characteristics of the fluid to be measured.
Electrode cover
Houses the electrodes. Do not remove the cover with the
detector installed on a pipe.
Union (Connected by
screws)
Hose
Clamp
The connection uses a union, hose, and clamp. The
material is SUS316.
Applicable for detector bore diameters of 2.5 to 15 mm.
Terminal box
(Remote model only)
Houses the connection terminals used to apply a standard
voltage.
Houses excitation and signal terminals.
Terminal box cover
(Remote model only)
Keeps the terminals box cover on during operation.