42
Cate-
gory
No.
Status messages
- ACTUAL_STATUS_NUMB
ER in the 'Advanced
Diagnostics' Transducer
Block .
- Local display
Error messages in the
Sensor Transducer Block
in question
Sensor
Transducer Block
measured value
status
Cause of error / remedy
Output
variables
affected
S-
502
Device status message (FF):
Linearization
S-502
Local display:
S-502 ↔ Measured value
BLOCK_ERR =
Other
|
Device needs
maintenance
|
Block
Configuration Error
QUALITY = BAD Cause of error:
Error in linearization.
Remedy:
Select valid type of linearization
(sensor type).
SV1, SV2,
PV1, PV2,
RJ1, RJ2
Transducer_Error =
Configuration error
SUBSTATUS =
Configuration
error
S-
901
Device status message (FF):
Ambient temperature too low
S-901
Local display:
S-901 ↔ Measured value
BLOCK_ERR =
QUALITY =
UNCERTAIN
Cause of error:
Reference junction temperature <
-40 °C (-40 °F); Alarm_Ambient_
Temp = OFF parameter.
Remedy:
Observe ambient temperature as per
specification.
SV1, SV2,
PV1, PV2,
RJ1, RJ2
Transducer_Error =
No error
SUBSTATUS =
Non-specific
F- 901
Device status message (FF):
Ambient temperature too low
F-901
Local display:
F-901
BLOCK_ERR =
Device needs
maintenance now
QUALITY = BAD Cause of error:
Reference junction temperature <
-40 °C ( -40 °F); Alarm_Ambient_
Temp = ON parameter
Remedy:
Observe ambient temperature as per
specification.
SV1, SV2,
PV1, PV2,
RJ1, RJ2
Transducer_Error =
General error
SUBSTATUS =
Device failure
S-
902
Device status message (FF):
Ambient temperature too
high
S-902
Local display:
S-902 ↔ Measured value
BLOCK_ERR =
QUALITY =
UNCERTAIN
Cause of error:
Reference junction temperature >
+85 °C (+185 °F); Alarm_Ambient_
Temp = OFF parameter
Remedy:
Observe ambient temperature as per
specification.
SV1, SV2,
PV1, PV2,
RJ1, RJ2
Transducer_Error =
No error
SUBSTATUS =
Non-specific
F-
902
Device status message (FF):
Ambient temperature too
high
F-902
Local display:
F-902
BLOCK_ERR =
Device needs
maintenance now
QUALITY = BAD Cause of error: Reference junction
temperature > +85 °C (+185 °F);
Alarm_Ambient_Temp = ON
parameter.
Remedy:
Observe ambient temperature as per
specification.
SV1, SV2,
PV1, PV2,
RJ1, RJ2
Transducer_Error =
General error
SUBSTATUS =
Device failure