Empty the pan.
Press key twice (double pressing the key helps to
avoid accidental starting calibration procedure).
During calibration internal weight is put three times on
and obtained results are compared.
Discrepancy of results is signalled with a message and
causes the balance being blocked.
Until calibration process is finished do not perform any
operation on the balance. Any vibrations and shocks
interfere calibration process and may delay it or
deteriorate accuracy of its result.
When internal calibration is performed successfully the
balance indicates zero on the display at empty pan.
Note:
In order to terminate calibration process press
CLR
key
and wait until balance mechanism is not settled in
initial position
AUTOCALIBRATION
Internal calibration:
(
- confirmation)
AUTOCALIBRATION
Please wait ...
AUTOCALIBRATION
Please wait ...
Max ... Min ... e= ... d= ...
0.0000
g
0%
100%