11 

Input Protection: 1000V dc or 1000V ac rms. 

 
Frequency (Auto-ranging) 

Range 

Resolution 

Accuracy 

4.000kHz 

1Hz 

+1.2% of rdg + 3 dgts 

40.00kHz 

10Hz 

400.0kHz 

100Hz 

10.00MHz 

1kHz 

+1.5% of rdg + 4 dgts 

Sensitivity: >0.5V RMS while ≤1MHz ; 
Sensitivity: >3V RMS while >1MHz ; 
Input Protection: 1000V dc or 1000V ac rms.  

Temperature 

Range 

Resolution 

Accuracy 

-20

o

C~+760

o

1

 o

+3% of rdg +5dgts 

-4

 o

F~+1400

 o

1

o

+3% of rdg +9dgts 

Sensor: Type K Thermocouple 

Overload protection: 1000V dc or ac rms.. 

Diode Test 

Test current 

Resolution 

Accuracy 

1mA typical/Open MAX.3V 

1 mV 

+10% of rdg + 5 digits 

Open circuit voltage: MAX. 3V dc  

Overload protection: 1000V dc or ac rms. 

Audible continuity 

Audible threshold:

 

Less than35Ω Test current MAX. 1.5mA 

Overload protection: 1000V dc or ac rms. 

ACCESSORIES 

Included accessories 

Standard Red/Black lead set with test probes 
 
 
 
 
 

Summary of Contents for AX-155

Page 1: ...TRUE RMS AUTORANGING DIGITAL MULTIMETER Model AX 155 INSTRUCTION MANUAL...

Page 2: ...ies recommend the use of fused test leads when measuring voltage on high energy systems Replacement fuses must be of the correct type and rating The instrument must not be used if any part of it is da...

Page 3: ...turn the range knob from the OFF position to any measurement range Note For best battery life ALWAYS turn the function switch to the OFF position when the meter is not in use This meter has Auto OFF...

Page 4: ...o normal AUTO measurement mode hold down the MAX MIN button for 2 seconds Peak Hold button The Peak Hold function captures the peak AC or DC voltage or current The meter can capture negative or positi...

Page 5: ...g into the negative COM jack 2 For current measurements up to 4000uA DC set the function switch to the uA position and insert the red test lead banana plug into the uA jack 3 For current measurements...

Page 6: ...til appears in the display 4 Touch the test probe tips across the circuit or part under test It is best to disconnect one side of the part under test so the rest of the circuit will not interfere with...

Page 7: ...citance measurement time can be several minutes before the final reading stabilizes The bar graph is disabled in capacitance measurement mode The LCD displays DIS C Discharging through the chip is qui...

Page 8: ...value Average responding multimeters are calibrated to read correctly only on sine waves and they will read inaccurately on non sine wave or distorted signals True rms meters read accurately on either...

Page 9: ...0mV 0 1mV 0 8 of rdg 3 digits 4 000V 1mV 40 00V 10mV 400 0V 100mV 1000V 1V 1 2 of rdg 5 digits Input Impedance 7 8M AC Response True RMS 50 Hz 60Hz Maximum Input 1000V dc or 1000V ac rms DC Current A...

Page 10: ...rms on mA 10A ac rms on 10A range Resistance Auto ranging Range Resolution Accuracy 400 0 0 1 0 8 of rdg 5 digits 4 000k 1 0 8 of rdg 2 digits 40 00k 10 400 0k 100 4 000M 1k 2 5 of rdg 8digits 40 00M...

Page 11: ...Resolution Accuracy 20 o C 760 o C 1 o C 3 of rdg 5dgts 4 o F 1400 o F 1 o F 3 of rdg 9dgts Sensor Type K Thermocouple Overload protection 1000V dc or ac rms Diode Test Test current Resolution Accurac...

Page 12: ...sing a Phillips head screwdriver 3 Insert the battery into battery holder observing the correct polarity 4 Put the battery cover back in place Secure with the two screws NOTE If your meter does not wo...

Reviews: