Product description
4.3
Description of the Gas Analysis System and Dimensions
The gas analysis system is located inside a stainless-steel wall cabinet. The space requirement and list
of the exterior and interior views are summarized in the table below. The cabinet is mounted to the wall
using four wall clips (see illustration for dimensions). There needs to be enough room in front of the
device for opening the door arrangement and for operating the device.
Table 1: Mounting Dimensions Gas Analysis System
Mounting dimensions [mm]
Device type
Width
Height
Depth
AwiFLEX XL
975
900
268
The analysis gas lines are routed to the system from the left hand side via the condensate traps for the
manual emptying of the system. Pressure-resistant pipes are recommended as analysis gas lines:
– Diameter 4/6mm or 1/8“/¼“
– Material: stainless steel 1.4401 or highly resistant plastic (depending on the site of operation,
stainless steel pipes might be obligatory)
All connections on the condensate traps and on the gas analysis system are implemented as hose screw
or pipe screw connections for these pipe cross-sections. For protection purposes, the analysis gas lines
should be run through protective tubes.
15