![Awite AwiECO Original Instruction Manual Download Page 23](http://html1.mh-extra.com/html/awite/awieco/awieco_original-instruction-manual_3035491023.webp)
Preparing the Gas Analysis System for Use
5.2.2.4
Connection cable for AI (analogue inputs) and AO (analogue
outputs)
Specifications Connection Cable:
•
AWG21 (metric equivalent 0.5mm²)
•
75°C / copper wire only (75°C / copper)
•
shielded / twisted in pairs
5.2.3
Potential equalization
The gas analysis system must be grounded. A screw is provided on the right-hand outer side of the
stainless steel housing for connecting the grounding cable. The grounding cable must have a cross section
of at least 6mm² (AWG 10).
WARNING
A non-connected grounding can lead to undefined voltages and electric shock. To unload voltages,
connect the potential equalization to the provided connections with the defined cross-section.
23