SPECIFICATIONS
Model:
AVR-E3-B
2
Maximum amplitude
3
: (50Ω load
8
)
100V
Rise time (20%-80%):
≤ 0.5 ns
Fall time (80%-20%):
≤ 1.0 ns
Pulse width (FWHM):
standard units: 10 ns to 200 ns
with -W1 option:
1 ns to 200 ns
with -W2 option: 50 ns to 5 us
with -W3 option:
1 ns to 5 us
Maximum PRF:
Standard units, or with -W1 option:
Units with -W2 or -W3 options:
100 kHz
20 kHz
Maximum duty cycle:
5%
Polarity:
Positive or negative or both (specify
4
)
DC offset or bias insertion:
Option available
5
. Apply required DC offset or bias in the range of ± 25 Volts,
(250 mA max) to back panel solder terminals.
Trigger modes:
Internal trigger, external trigger (TTL level pulse, > 10 ns, 1 kΩ input impedance),
front-panel “Single Pulse” pushbutton, or single pulse trigger via computer command.
Variable delay:
0 to 1.0 seconds, for all trigger modes (including external trigger).
Propagation delay:
≤ 160 ns (Ext trig in to pulse out)
Jitter: (Ext trig in to pulse out)
± 35ps ± 0.015% of sync delay
Sync output:
> +3 Volts, > 50 ns, will drive 50 Ohm loads
Gate input:
Synchronous or asynchronous, active high or low, switchable. Suppresses triggering when active.
Connectors:
Out: SMA, Other: BNC
Power requirements:
100 - 240 Volts, 50 - 60 Hz
GPIB and RS-232 control
2
:
Standard on -B units.
LabView drivers:
Check
http://www.avtechpulse.com/labview
for availability and downloads
Ethernet port, for remote control
using VXI-11.3, ssh, telnet, & web:
Optional
6
. Recommended as a modern alternative to GPIB / RS-232.
See
http://www.avtechpulse.com/options/vxi
for details.
Settings resolution:
The resolution of the timing parameters (pulse width, delay, period) varies,
but is always better than 0.15% of (|set value| + 20 ns).
The amplitude resolution is < 0.1% of the maximum amplitude.
Settings accuracy:
Typically ± 3% (±2 ns or ± 2% of max. amplitude) after 10 minute warmup. For high-accuracy
applications requiring traceable calibration, verify the output with a calibrated oscilloscope.
Dimensions (H x W x D):
100 mm x 430 mm x 375 mm (3.9” x 17” x 14.8”)
Chassis material:
cast aluminum frame and handles, blue vinyl on aluminum cover plates
Temperature range:
+5°C to +40°C
1)
-C suffix indicates stand-alone lab instrument with internal clock and line powering. (See
http://www.avtechpulse.com/formats
for additional details of
the four basic instrument formats).
2)
-B suffix indicates IEEE-488.2 GPIB and RS-232 control of amplitude, pulse width, PRF and delay (see
http://www.avtechpulse.com/gpib
).
3) For operation at amplitudes of less than 10% of full-scale, best results will
be obtained by setting the amplitude near full-scale and using external attenuators on the output.
4) Indicate desired polarity by suffixing model number with -P or -N (i.e. positive or negative) or -PN for dual polarity option. Polarity reversal achieved
by means of a two-position switch that controls the polarity of the signal output port on -C units, and via keypad control on -B units. -PN option not
available on -W3 units.
5) For DC offset option suffix model number with -OS.
6) Add the suffix -VXI to the model number to specify the Ethernet port..
7) Fall time increases to < 3 ns for pulse widths less than 15 ns.
8)
A 50Ω load is required. Other loads may damage the instrument. Consult Avtech (
) if you need to drive other
load impedances.