background image

INSTRUCTIONS

MODEL AVMH-4-C

100 VOLTS, 10 MHz 

IMPULSE GENERATOR 

WITH 1 ns PULSE WIDTH

SERIAL NUMBER: ____________

A V T E C H     E L E C T R O S Y S T E M S     L T D .

N A N O S E C O N D   W A V E F O R M   E L E C T R O N I C S

S I N C E   1 9 7 5

P.O. BOX 265

OGDENSBURG,  NY

U.S.A. 13669-0265

TEL: 888-670-8729 (USA & Canada) or +1-613-686-6675 (Intl)

FAX: 800-561-1970 (USA & Canada) or +1-613-686-6679 (Intl)

[email protected]    -    http://www.avtechpulse.com/

X BOX 5120, LCD MERIVALE

OTTAWA, ONTARIO

CANADA    K2C 3H5

Summary of Contents for AVMH-4

Page 1: ...Y S T E M S L T D N A N O S E C O N D W A V E F O R M E L E C T R O N I C S S I N C E 1 9 7 5 P O BOX 265 OGDENSBURG NY U S A 13669 0265 TEL 888 670 8729 USA Canada or 1 613 686 6675 Intl FAX 800 561...

Page 2: ...ace said defective item This warranty does not apply to units which have been dissembled modified or subjected to conditions exceeding the applicable specifications or ratings This warranty is the ext...

Page 3: ...INSTALLATION 10 VISUAL CHECK 10 POWER RATINGS 10 CONNECTION TO THE POWER SUPPLY 10 PROTECTION FROM ELECTRIC SHOCK 11 ENVIRONMENTAL CONDITIONS 12 FUSES 13 AC FUSE REPLACEMENT 13 DC FUSE REPLACEMENT 14...

Page 4: ...OW VOLTAGE POWER SUPPLY 1 3 26 PCB 158P LOW VOLTAGE POWER SUPPLY 2 3 27 PCB 158P LOW VOLTAGE POWER SUPPLY 3 3 28 PCB 235A HIGH VOLTAGE DC POWER SUPPLY 29 PCB 219E OSCILLATOR AND TRIGGER CIRCUIT 30 PCB...

Page 5: ...g transformer that can be installed on the output to generate a negative signal Instruments with the N PN suffix generate up to 100V at the main output and are supplied with an inverting transformer t...

Page 6: ...e http www avtechpulse com formats for additional details of the basic instrument formats 2 Indicate desired polarity by suffixing model number with P or N i e positive or negative or P PN or N PN for...

Page 7: ...g Also check that your load is adequately shielded It may be necessary to enclose the load in a metal enclosure If any of the connectors on the instrument are unused they should be covered with shield...

Page 8: ...ies to Directive 2002 96 EC categories 1 7 and 10 as stated in the Article 2 Scope section of Directive 2002 95 EC DIRECTIVE 2002 96 EC WEEE European customers who have purchased this equipment direct...

Page 9: ...ura AAD130SD 60 A switching power supply According to the manufacturer the Tamura AAD130SD 60 A has the following certifications UL60950 1 IEC60950 1 CSA C22 2 No 60950 1 EN60950 1 and is compliant wi...

Page 10: ...NNECTION TO THE POWER SUPPLY An IEC 320 three pronged recessed male socket is provided on the back panel for AC power connection to the instrument One end of the detachable power cord that is supplied...

Page 11: ...ors are prevented access and or are insulated from every connection point In some cases connections must be exposed to potential human contact Operators must be trained to protect themselves from the...

Page 12: ...ndoor use 2 altitude up to 2 000 m 3 temperature 5 C to 40 C 4 maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C 5 Mains supply voltage fluct...

Page 13: ...the fuses on the rear panel are shown in the figure below AC FUSE REPLACEMENT To physically access the AC fuses the power cord must be detached from the rear panel of the instrument The fuse drawer ma...

Page 14: ...inal Mains Voltage Rating Case Size Recommended Replacement Part Littelfuse Part Number Digi Key Stock Number 1 2 AC 100 240V 0 5A 250V Time Delay 5 20 mm 0218 500HXP F2416 ND 3 DC N A 1 6A 250V Time...

Page 15: ...ved the instrument will resume normal operation This overload indicator may flash yellow briefly at start up This is not a cause for concern 3 PRF Range Switch and Fine Dial This switch sets the pulse...

Page 16: ...the SYNC pulse width varies but is always 10 ns When triggering externally the SYNC output is a buffered replica of the TRIG input The amplitude is fixed at 3V approximately and this output will driv...

Page 17: ...NPUT CONNECTOR Present on units with os option only A DC offset in the range of 50V 250 mA max may be applied to this solder terminal The DC offset will appear on the output When this feature is not u...

Page 18: ...PULSE CONTROL This instrument can be triggered by its own internal clock or by an external TTL trigger signal When triggered internally two mainframe output channels respond to the trigger OUT and SY...

Page 19: ...POLARITY INVERSION Instruments with the P PN suffix generate up to 100V at the main output and are supplied with an AVX 2 inverting transformer that can be installed on the mainframe output A negative...

Page 20: ...added to the DC terminal of the inverting transformer Do not apply the offset to the rear panel offset terminal on the mainframe if present PULSE STABILITY If the output pulse appears to be unstable...

Page 21: ...here L is the inductance ILOAD is the load current change and t is time For this reason it is important to keep any parasitic in the load low This means keeping wiring short and using low inductance c...

Page 22: ...ernal offset must be added to the DC terminal of the inverting transformer Do not apply the offset to the rear panel offset terminal on the mainframe if present M OPTION This option provides a monitor...

Page 23: ...contact Avtech info avtechpulse com to arrange for the instrument to be returned to the factory for repair Service is to be performed solely by qualified service personnel Caution High voltages are pr...

Page 24: ...es may require replacement All fuses can be accessed from the rear panel See the FUSES section for details CLEANING If desired the interior of the instrument may be cleaned using compressed air to dis...

Page 25: ...h a s s is g ro u n d p o s t G 1 G 2 B 1 R E D G 4 G 3 L N G X 2 C O R C O M 6 E G G 1 2 P O W E R E N T R Y M O D U L E 1 1 b 1 a 2 2 b 2 a X 1 P O W E R S W I T C H S W 3 2 5 N D C W IN D U S T R...

Page 26: ...N D 1 5 V P O U T 3 P O U T 1 1 5 V 1 2 J 7 6 4 0 4 5 6 2 D ig i A 1 9 2 1 N D X 8 6 3 2 1 4 S S S C R E W 0 6 0 4 M P P 1 8 8 X 9 6 3 2 1 4 S S S C R E W 0 6 0 4 M P P 1 8 8 X 1 3 6 3 2 S S E X T T...

Page 27: ...ig i P 5 5 5 0 N D C 8 2 2 u F D i g i 4 4 5 2 8 9 6 N D L 1 M o u s e r 4 3 4 7 7 A 1 0 1 M 0 1 C 4 4 7 u F 3 5 V D i g i P 5 5 5 0 N D C 1 2 2 2 u F D i g i 4 4 5 2 8 9 6 N D L 4 M o u s e r 4 3 4 7...

Page 28: ...A V T E C H D I S A B L E O L O W H E N C H A R G I N G S H O R T S O U T B A S E W H E N C H A R G I N G B U 1 5 V G N D 1 5 V X 1 K E Y S T O N E 6 2 1 D i g i 6 2 1 K N D R 1 5 0 I F O L O U S E S...

Page 29: ...PCB 235A HIGH VOLTAGE DC POWER SUPPLY...

Page 30: ...PCB 219E OSCILLATOR AND TRIGGER CIRCUIT...

Page 31: ...PCB 206B PRF TO VOLTAGE...

Page 32: ...PCB 216D PRF PW TWEAKING PCB...

Page 33: ...I O N C C W C W W R 3 T R 1 K C C W C W W R 1 A M P 1 K 1 5 V G N D 1 5 V G N D T R I G O U T S M A G N D S Y N C 5 V L O F 5 V H I F B I A S M 2 C L 8 2 1 9 E C O N N 3 T R IG C O N N 4 S Y N C 5 V 1...

Page 34: ...PERFORMANCE CHECK SHEET 34...