4000
4-/5-Gas Measuring Instruments Unit
Service Manual
5-3
5.2.1 Light
Source
To carry out repair work:
•
Disconnect the inlet and outlet hose from the measuring cell.
•
Remove the complete gas analyzer from the base plate (see Dismantling the 4-/5-Gas
Measuring Instruments Unit).
•
Undo the 4 screws on the IR unit.
•
Unsolder the two red cables to the IR unit from P3.
•
Then pull the IR unit out of the housing.
•
Mount the new IR unit.
•
Lay the litz wires underneath the main board and solder them into P3/1,2.
•
Turn the analyzer back to its proper position and screw it to the base plate.
•
Connect the inlet hose to the IR side of the measuring cell.
•
Connect the outlet hose to the detector side of the measuring cell.
5.2.2 Motor
Unit
•
Disconnect the inlet and outlet hoses from the measuring cell.
•
Remove the complete gas analyzer from the base plate (see Dismantling the 4-/5-Gas
Measuring Instruments Unit)
•
Disconnect the 4-pin plug (P5) from the main board.
•
Remove the motor unit by undoing 2 screws.
•
Mount the new unit (reassemble in reverse order).
Never touch the ceramic and aluminium parts with your
fingers!
Never clean quartz glass with alcohol. Only clean the quartz
glass plates when you change the measuring cell.
Summary of Contents for AVL DiCom 4000
Page 4: ......
Page 6: ......
Page 22: ...Periodic Maintenance AVL 4000 1 12 Service Manual ...
Page 47: ...4000 Evaluation Unit Service Manual 4 15 4 9 Main Board 4 9 1 Components Location Diagram ...
Page 48: ...Evaluation Unit AVL 4000 4 16 Service Manual 4 9 2 Circuit Diagrams ...
Page 49: ...4000 Evaluation Unit Service Manual 4 17 ...
Page 50: ...Evaluation Unit AVL 4000 4 18 Service Manual ...
Page 51: ...4000 Evaluation Unit Service Manual 4 19 ...
Page 52: ...Evaluation Unit AVL 4000 4 20 Service Manual ...
Page 54: ...Evaluation Unit AVL 4000 4 22 Service Manual 4 10 2 Circuit Diagrams ...
Page 55: ...4000 Evaluation Unit Service Manual 4 23 ...
Page 56: ...Evaluation Unit AVL 4000 4 24 Service Manual ...
Page 57: ...4000 Evaluation Unit Service Manual 4 25 ...
Page 59: ...4000 Evaluation Unit Service Manual 4 27 4 11 2 Circuit Diagram ...
Page 61: ...4000 Evaluation Unit Service Manual 4 29 4 12 2 Circuit Diagram ...
Page 63: ...4000 Evaluation Unit Service Manual 4 31 4 13 2 Circuit Diagram ...
Page 65: ...4000 Evaluation Unit Service Manual 4 33 4 14 2 Circuit Diagram ...
Page 67: ...4000 Evaluation Unit Service Manual 4 35 4 15 2 Circuit Diagram ...
Page 69: ...4000 Evaluation Unit Service Manual 4 37 4 16 2 Circuit Diagram ...
Page 71: ...4000 Evaluation Unit Service Manual 4 39 4 17 2 Circuit Diagram ...
Page 73: ...4000 Evaluation Unit Service Manual 4 41 4 18 2 Circuit Diagram ...
Page 75: ...4000 Evaluation Unit Service Manual 4 43 4 19 2 Circuit Diagram ...
Page 77: ...4000 Evaluation Unit Service Manual 4 45 4 20 2 Circuit Diagram ...
Page 79: ...4000 Evaluation Unit Service Manual 4 47 4 21 2 Circuit Diagram ...
Page 124: ...4 5 Gas Measuring Instruments Unit AVL 4000 5 28 Service Manual ...
Page 149: ...4000 Opacity Measuring Chamber Type A Service Manual 6 25 Fig 6 30 Fig 6 31 ...
Page 150: ...Opacity Measuring Chamber Type A AVL 4000 6 26 Service Manual Fig 6 32 ...
Page 180: ...Spare Parts Lists AVL 4000 8 14 Service Manual ...
Page 182: ...Tools for the Service Technician AVL 4000 9 2 Service Manual ...