Maintenance
24
AVL 415SE Smoke Meter — Maintenance Guide
1.
Carry out a reference measurement with a "new" line, if it seems to be nec-
essary.
2.
Replace the sampling lines (Viton hoses) regularly.
Recommended replacement intervals: every 1 000 … 2 000 operating hours.
The material number depends on the length of the sampling line
2.3.5
Suction Unit
Clamp piece
tracks
1.
Check the friction conditions in the clamp piece tracks.
2.
Remove the hose from the upper clamp piece
3.
Release the clamping lever and open it completely.
4.
Check the clamp piece tracks for easy movement.
5.
If necessary, lubricate track (ex-works: Teflon lubricant spray - part number
HS0058, Teflon lubricant oil - part number HS0064, or Lubricant grease
OKS 260 - part number HS0071)
Fig. 7 Sampling line
Information
Heavier soot deposits can occur when using new sampling lines for the first time
straight from the factory and this can also reduce the recorded FSN values.
The lower the measurement values, the higher, the requirement for repeatability,
and reproducibility.
The greater is the influence of the soiling along the entire sampling path.
Fig. 8 View of the suction unit in the device
Information
Remember the hose clip when disconnecting and reconnecting the hose.