AVIV 410 CD Standard Operating Procedure Download Page 1

Standard operating procedure for Aviv 410 CD instrument 

 

Turning ON the instrument: 

1.

 

Turn on the N

2

 gas bottle by opening the gas valve on top (1/4 turn) and 

not the regulator (do not change the settings on the regulator). 
Familiarise yourself with the gauge: a full bottle reads 200 Litres, as 
measured on the inner red dial. 

2.

 

Open the inlet and outlet valves on both sides of the oxygen scrubber on 
the floor next to the CD instrument. 

3.

 

Adjust the N

2

 flow regulator to 25. 

4.

 

Turn on the water-bath pump situated on the unit to the right of the CD 
instrument. The temperature of the water bath is set to 20˚C. Do not 
change this setting. Check the water level and add de-ionised water if 
necessary. Make sure that the water is circulating. 

5.

 

Monitor the oxygen level on the oxygen monitor on the right hand side 
of the lower shelf of the CD trolley. The oxygen level should read below 
5 ppm before you can operate the CD instrument. Adjust the flow rate in 
the oxygen monitor by turning the black knob at the back of the monitor 
clockwise or anti-clockwise. The black marker in the flow meter on the 
front of the oxygen monitor should always be ~ 0.2 (otherwise you are 
not measuring the true oxygen concentration). Make sure that the 
instrument is reading ppm and not % (it changes automatically when it 
reaches 0.1% oxygen (1000 ppm) to the ppm scale). There are indicator 
lights next to the displayed number that tell you if the units are ppm or 
percentage. This step should take about 60-90 min. Do not attempt to 
turn-on the xenon lamp before the read-out is below 5 ppm.  

6.

 

Turn-on the switch marked ‘Xenon lamp’ on the front left side of the 
CD instrument. 

7.

 

The lamp power supply is on the lower shelf next to the oxygen 
monitor. Wait until the LED green light, indicating ‘lamp ready’, is on. 
When the green light is on, press the red button on the lamp power 
supply to ignite the lamp. 

8.

 

Turn on the CPU switch on the front of the instrument (next to the 
Xenon lamp switch). 

9.

 

Turn on the computer and open the Aviv 410 software. 

10.

 

Watch the oxygen level in the oxygen monitor and gradually reduce the 
N

2

 flow to 20 while at the same time maintaining the flow rate in the 

oxygen monitor to ~0.2 (this could be done at any time during the 
experimental run). 

You are now ready to start the experiments. 

11.

 

During operation, monitor the N

2

 purity by reading the output of the 

oxygen monitor. If at any point the oxygen level goes above 5 ppm, turn 
the Xenon lamp off immediately. As well as the oxygen level you 

Reviews: