11
EMC Tables
NE10
is intended for use in the electromagnetic environment specified below. The
customer or the user of
NE10
must make sure that it is used in such an
environment.
Guidance and manufacturer’s declaration - Electromagnetic emissions
Phenomenon
Professional healthcare
facility environment a)
HOME HEALTHCARE
ENVIRONMENT a)
Conducted and radiated
RF MISSIONS
a)
CISPR 11
Group 1 Class B
Harmonic distortion
Not applicable
Voltage fluctuations and
flickering
Not applicable
a) The equipment is suitable for use in Home Health Environments and
Professional Health Care Environments limited to patient rooms and respiratory
treatment facilities in hospital or clinics. The more restrictive acceptance limits of
Group 1 Class B (CISPR 11) have been considered and applied. The equipment is
suitable for use in the mentioned environments when directly connected to the
Public Mains Network.
b) The test is not applicable in this environment unless the ME EQUIPMENT and
ME SYSTEM used will be connected to the PUBLIC MAINS NETWORK and the
power input is otherwise within the scope of the Basic EMC standard.
Guidance and manufacturer’s declaration - Electromagnetic immunity - Enclosure
port
Phenomenon
Basic EMC
standard or test
method
Immunity test levels
Professional
healthcare
facility
environment
HOME HEALTHCARE
ENVIRONMENT
ELECTROSTATIC
DISCHARGE
IEC 61000-4-2
± 8kV contact
± 2 kV, ±4kV ±, ±8 kV, ±15 kV air
Radiated RF EM
fields
IEC 61000-4-3
a)
10 V/m b) 80MHz - 2.7 GHz
80% AM at 1kHz
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
COMPLIANT
NOTE: Further information about distances
to be maintained between portable and
mobile RF communications equipment
(transmitters) and the
NE10
can be
requested from supplier using the contact
information provided in this manual.