66
ZM400 Series Indicators Service Manual
6
ADMIN level menus
6.5.1 Serial
SEriAL
Use this to set the serial communication parameters. Choose Port 1or Port
2 to begin.
bAud
Set the baud rate from the available choices.
9600
is the
default value.
d-bitS
Choose
7
or
8
data bits.
8
is the default.
PAritY
Choose
nonE
,
odd
or
EVEn
.
nonE
is the default.
S-bitS
Choose
1
or
2
stop bits.
1
is the default.
c-troL
Use this to set the flow control for the port. Choices for Port
1 are
nonE
,
hArd
or
SoFt
. Choices for Port 2 are
nonE
or
SoFt
. Default is
nonE
for both ports.
hArd
Stands for hardware flow control. If wired, a
device can stop the indicator and restart it for
transmitting. Example is the CTS signal on a PC
serial port or print busy output on a Line or Lister
printer.
SoFt
Stands for software handshaking. This allows a
device to start and stop the indicator transmit by
sending Xon or Xoff characters.
Port 1 has both
hArd
and
SoFt
. Port 2 does not
have hardware handshaking.
Port1
Port2
Baud
D-Bits
Parity
S-Bits
C-trol
Select
300
600
7
8
None
Odd
Even
1200
2400
4800
9600
19200
38400
57600
115200
1
2
None
Hard
Soft
Port1-2
Port1-H
Serial
Port 1
only
Port 1
only
Hardware flow control on Port 1 is only available if Port1-H is selected and Jumper P5
is in position 1.
See the photo in section