ZM400 Series Indicators Service Manual
43
6.2
Calibration Procedure
Pt2-Pt9
The points are numbered 2 through 9 because, internally, the
zero reference point is point 1 and the span point is point 10.
Follow the same steps as described in the SPAN procedure to
enter each linearity point.
6.2.5 Input procedure
ZEro
Use this to enter a value for the zero point.
CountS
Use this to enter a zero point using ADC counts.
mV
Use this to enter a zero point using a mV/V value.
SPan
Accept the flashing displayed span weight (
XXXX
) or key in the span
weight that corresponds with the span ADC or mV/V value.
CountS
Use this to enter a span using ADC counts.
mV
Use this to enter a span using a mV/V value.
Linearity points are cleared if a new span calibration is performed.
Input
Zero
Span
Counts
mV
Counts
mV
XXXX
The BSQ base only allows entry of Counts for Zero or Span points
The Span value is the differential value of the actual Calibration Zero and Span count
(or mV/V) values.