39
2.3 C
ONNECTING
THE
SENSORS
ALS
306
309
Adjusting the Wenglor photoelectric
sensor automatically
Only for Wenglor sensors (old standard, replaced since
9/2011 by Avery sensor).
Call up
MACHINE SETUP > Auto Sensor Adj.
– The label material is fed slowly until the next gap
between labels (punch) is reached. The photoelec-
tric label sensor is automatically adjusted in the
process.
Adjusting the Wenglor photoelectric
sensor manually
Only for Wenglor sensors (old standard, replaced since
9/2011 by Avery sensor).
For dispensing label material with a low contrast,
it can be necessary to adjust the label sensor
manually:
Call up
MACHINE SETUP > Sensor Adjust
.
– Display:
Place backing paper in the fork of the photoelectric
sensor.
Press the
key, until the LED [68A] on the
photoelectric sensor lights up.
Press the
key and keep it pressed just beyond
the point where the LED goes out.
Note the displayed value (value 1).
Place label material in the fork of the photoelectric
sensor.
– The LED will light back up.
Press the
key and keep it pressed just beyond
the point where the LED goes out.
Note the displayed value (value 2).
Calculate value 3:
– Value 3= ( value 1 + (value 2 - value 1) / 2)
Set value 3 by means of the two arrow keys and
acknowledge the adjustment by pressing the
key.
Sensor Adjust
Manual:
xx%
[68] LED (A) at the (Wenglor) photoelectric label sensor
A