7.3.7 Set Pressure Control Level
1) Select the [Pinsp] hotkey.
2) Press the control knob and turn the knob to set Pinsp at the desired value.
3) Press the control knob to confirm and activate the change.
7.3.8 Set Pressure Support Level
1) Select the [Psupp] hotkey.
2) Press the control knob and turn to set [Psupp] at the desired value.
3) Press the control knob to confirm and activate the change.
7.3.9 Set Apnea Time
If SPONT / PSV mode is used, the system is provided with a backup ventilation mode.
If an Apnea situation occurs, and there is no spontaneous breathing, or spontaneous breathing
and an inspiratory trigger condition is not reached, the system in accordance with the set ‘Apnea
time', automatically enters backup ventilation mode.
Apnea Time in a no-breath conditions is set as follows:
1) Select [Alarm] -> [Apnea time].
2) Turn the control knob to set the Apnea time to the desired value.
3) Press the button to confirm and activate the change.
4) Closing the window or returning to the main screen will cancel the setting.
7.3.10 Set Inspiratory Pause
1) In VCV mode, select the [Pause] hotkey.
2) Set the pause at the desired value:0, 5%, 10%, 15%, 20%, 25%, 30%, 35%, 40%, 45%, 50%,
55%, 60%.
3) Press the control knob to confirm and activate the change.
7.3.11 Set Rise Time
In pressure mode, rise time is available.
1) Select [Tslope] hotkey.
2) Set the [Tslope] at the desired value, namely 0-2s, 0.1s increments.
3) Press the control knob again to activate the change.
7.3.12 Set Expiratory trigger sensibility
1) In the SIMV-V, SIMV-P, SPONT / PSV mode, select [Control] - [Esens] hotkey.
2) With the control knob pressed, turning the knob to set Esens to the desired value, (5% to 80%
with increments of 5%).
3) Press the control knob to confirm the change.