24
6.2
Cleaning electrode of the probe
TURBOXY
After overchlorination with the TURBOXY chlorine probe in place, or a prolonged shutdown
of several days, the electrode can be oxidized and no longer allow the chlorine level to be
measured correctly.
Remove the mask and the turbine of the probe.
Clean the electrode with some sandpaper P500 see smaller. The purpose is to
remove the corrosion of the electrode (axis of the turbine).
Install the mask and the turbine on the probe TURBOXY
a) Assembly view
b) Slide the turbine on the axis
c) Locate the spur
d) Slide the mask with the
spur in the notch
Suivre la procédure de calibrage de la sonde chlore TURBOXY cf : « Calibration of the Turboxy
chlorine probe. »
6.3
Maintenance of the pH electrode
6.3.1
Care instructions for the pH electrode
Over time, in addition to the normal wear and tear of the electrode, depending on the use and quality of
the water, the measurement will deteriorate. A thin layer of limescale, together with other elements
present in the pool water, will be deposited on the sensitive elements of the sensor. Pour palier à ce
problème, il est conseillé d’utiliser la solution de nettoyage pour électrode pH ou redox AYACSOLNET02 et
de suivre la procédure indiquée avec le flacon. To overcome this problem, it is recommended to use the
cleaning solution for pH or redox electrodes AYACSOLNET02 and to follow the procedure indicated on the
bottle. We advise you to change the electrode, f the measurements no longer give good results or the
quality displayed after calibration is lower than 50, and changing the standard solution if you have any
doubts about its quality after several uses or after the expiry date.