SAVe II+ OPERATOR’S MANUAL > Page 23
S
TEP
3:
C
ONNECT
P
ATIENT
B
REATHING
C
IRCUIT
Locate an unused SAVe II+ Patient Breathing Circuit in its original packaging. Remove
patient circuit from packaging and connect to the appropriate SAVe II+ Manifold Port.
S
TEP
4:
T
URN ON
.
S
ELECT PATIENT HEIGHT
.
P
RESS CONFIRM
.
The Adult Height Presets
default to 6 ml/kg of ideal body
weight for a male. There is a
look up chart in the Quick
Start Guide and on the bottom
of the device that lists the tidal
volume for male and females
which can be used to make
adjustments. At 6 ml/kg of
ideal body weight females
receive 30 ml less volume
than males of the same
height. Do not forget to add
PEEP if it is indicated.
Figure 16: Breathing Circuit Connected
Figure 17: Setup Using Adult Height Presets
Figure 15: Air/ O
2
Intake Port and Breathing Circuit Connection Ports