
PAGE 12
HSP SERIES INSTALLATION AND OPERATION MANUAL
SOFTWARE SETUP (CONT)
Fault
The FAULT output will be set low (0V) under the following conditions:
•
The amplifier has gone into thermal shutdown due to a measured high temperature
•
The amplifier has detected an overload condition on the speaker terminals
The FAULT output is located on the rear of the HSP unit. Connect the FAULT and GND connection to any external control solution to read this signal.
The FAULT output is an open collector output with a 390R current limiting resistor in series.
Maximum voltage relative to GND
35V
Maximum current
100mA
Priority 1 – External Input
The standby function is activated using the supplied 4-way Phoenix connector.
Simply connect the STANDBY to GND to put the amplifier into standby mode.
See connection diagram listed below:
FAUL
T
GND
ST
ANDBY
MUTE
Priority 2 – Command Interface
Send the following command using RS232
To enter standby:
set device mode standby true
To exit standby:
set device mode standby false
You can also enable/disable the software standby using the PC application.
Open the standby dialog box by selecting the “Standby” button and then select
the Standby button under the priority 2 section.
Priority 3 – Auto standby
The amplifier can enter and exit standby based on the presence of audio
on any input. (Note: All input channels are summed together for detection)
Open the standby dialog box in the PC application by selecting the “Standby”
button and set the time and audio threshold under the priority 3 section.
FAULT
390R
GND