User and Technical Manual: HotDog Temperature Management Controller
Page 20
Alarm Error Condition
Code
Condition
Delay
Signal Generation Delay
(Software Alarm)
Signal Generation
Delay
(Hardware Alarm)
Overcurrent (System) [hardware]
E3
< 1 millisecond
< 200 milliseconds
< 50 seconds
Primary Over-temperature [software]
E1
15 Sec
< 200 milliseconds
(software only;
won’t alarm)
Port Current Limit Reached [hardware]
E3
< 1 millisecond
< 200 milliseconds
< 50 seconds
Sensor or Cable Failure [software]
E4
15 Sec
< 200 milliseconds
(software only;
won’t alarm)
Blanket Fold Detection [software]
E5
15 Sec
< 200 milliseconds
(software only;
won’t alarm)
Over Temperature TruCore (NA) [hardware]
E6
< 1 millisecond
Not implemented
< 50 seconds
Over Temperature (Secondary) [hardware]
E8
< 1 millisecond
< 200 milliseconds
< 50 seconds
SPECIFICATIONS
Physical Characteristics
Dimensions
28 cm high x 17.8 cm deep x 22.2 cm wide (11
” high x
7
” deep x
8
.75” wide
)
Weight
2.8-3.2 kg (6.25
–
7.15 lbs.) without the clamp
Mounting
Can be placed on a horizontal flat surface (i.e. table top), clamped to an IV pole, or hung using a VESA mount of
either FDMI MIS-C (35 × 75 mm) or FDMI MIS-D (75 × 75 mm) specifications
Temperature Characteristics
Temperature Control
Micro-processor
Operating Temperatures
Blanket Ports A, B, C, and D adjustable in 1
o
C increments
37° to 43° ± 1.0°C
98.6° to 109.4° ± 1.8°F
Mattress Port M adjustable in 1
o
C increments
35° to 40° ± 1.0°C
95° to 104° ± 1.8°F
Safety System
All alarm conditions are classified as Medium Priority Technical Alarms
Auditory Alarms
Minimum SPL of 65 dB(A) at 3m (from front of controller) with a back ground SPL not to exceed 55dB(A)
Primary Over-temp Alarm
Ports A, B, C, D (Warming Blanket)
Medium Priority Alarm sounds when temperature sensor is at set point + 1°C
Port M (Warming Mattress)
Medium Priority Alarm sounds when temperature sensor at set point + 1°C
Secondary Over-temp Alarm
Ports A, B, C, D (Warming Blanket)
Independent electronic circuit shuts the heater off if the Warming Blanket temperature sensor reaches max set