cryo
ICE BOX Error Codes
If an error condition should occur, the Maintenance Needed Indicator or the System Fault
Indicator will illuminate. The Probe Temperature display on the front panel will
temporarily display one of the following error codes during the power-up sequence.
Contact AtriCure Service if one of these conditions occurs.
Error ID
Error
Likely Cause
001
No 24 VDC
Fuse (F2)
002
Cylinder over temperature
Heater blanket
003
Probe overpressure
Pressure regulator
004
Unwanted Probe pressure
Leaky inlet valve
005
No 230 VAC
Fuse (F1)
008
Cylinder over pressure/temperature
Overheated Cylinder
PPP
Power On Self-Test Error
Activation Button/Footswitch
Pressed during power up
10.
ELECTROMAGNETIC COMPATABILITY TABLES
Electromagnetic Emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
The
cryo
ICE
BOX
is intended for use in the electromagnetic environment specified below. The customer or the
user of the
cryo
ICE
BOX
unit should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The
cryo
ICE
BOX
unit uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
The
cryo
ICE
BOX
unit is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Complies
xxxvii