TROUBLESHOOTING
Problem
Possible Cause
Action
Front displays
not lit
No power
Check power switch on back
of
cryo
ICE
BOX
Check plug connection on
back of
cryo
ICE
BOX
Check AC plug in wall socket
Ensure power is available at
wall socket
cryo
ICE
BOX electrical failure
Call AtriCure Service
Cylinder Heater
Band Icon
Illuminated
Heater not plugged in
Check connection on back of
unit
N
2
O cylinder valve closed
Ensure N
2
O valve is open
Empty N
2
O cylinder
Replace N
2
O cylinder
Extremely cold N
2
O cylinder
Allow 15 minutes to warm up
Heater not attached to N
2
O
cylinder
Attach Heater Band to
cylinder
Heater malfunctioning
Call AtriCure Service
Temperature
Not Displayed
cryo
ICE Probe not plugged in
Ensure
cryo
ICE Probe
thermocouple leads are
firmly seated within their
receptacles
Malfunctioning
cryo
ICE Probe
Replace
cryo
ICE Probe
cryo
ICE
BOX malfunctioning
Call AtriCure Service
cryo
ICE
BOX has
power but will
not go into
FREEZE mode
cryo
ICE Probe not plugged in
Plug in
cryo
ICE Probe
N
2
O cylinder empty
Replace N
2
O Cylinder
N
2
O cylinder valve closed
Open cylinder valve
Inlet Gas Connection not
secure
Ensure Inlet Gas Connection
is completely seated