Calibration
7.4 Calibration- Assigning/changing names
Assign a unique name to the calibration (calibration method, material desig-
nation, batch no., …), example: Cal-1Foil EN AW 6082
Keep in mind that many calibrations with different calibration methods are
stored in the gage. A unique name makes it easier to select and assign
the desired calibration to a batch.
Procedure
1. Open the
Rename
function:
Main Menu (
) > Calibrations >
Name of cali-
bration
> Rename
2. Change name:
•
Move cursor by means of the
BACK
and
OK
keys
•
Select the desired character by means of the keys
•
Delete character: Select space
3. Exit the
Rename
input window:
a Move the cursor entirely to the right using the
OK
key.
The character
appears at the right edge of the display.
b Press
OK
to exit the input window.
7.5 Calibration - Reset
The reset function ensures that all coating thickness correction values for the
selected calibration are deleted. The
Reset
function deletes all parameters mea-
sured in the foil calibration step; the nominal values for the foil are retained,
however. Measurements in the zero calibration step are retained.
Procedure
1. Open the
Reset
function:
Main Menu (
) > Calibrations >
Name of cali-
bration
> Reset > OK
2. Prompt asking whether you really wish to perform reset, confirm with
YES
.