ATO HADV-1T Instruction Manual Download Page 1

Series

 

Touch

 

Screen

 

Viscometer

Instruction Manual

Welcome to visit our website! Any questions, 

please contact us...

>> 

Email: [email protected]

>> Phone: +1 800-585-1519

>> Website: www.ato.com

Summary of Contents for HADV-1T

Page 1: ...T Series Touch Screen Viscometer Instruction Manual Welcome to visit our website Any questions please contact us Email sales ato com Phone 1 800 585 1519 Website www ato com...

Page 2: ...through this Instruction Manual before correctly using this product Any further revision will not be informed separately Contents I Working principle and purpose 3 II Main technical performance 3 III...

Page 3: ...tor and a combination of rotating speeds directly fast and accurately save the identified testing conditions and use them conveniently for future testing This series of instrument can be extensively u...

Page 4: ...the base lifting pole viscometer main machine main machine connecting bar and power supply adapter etc As shown in Fig 1 insert the lifting pole into the base hole and tighten by the nut notice the l...

Page 5: ...tor protective frame is equipped for series HA HB viscometers Fig 2 Fig 3 7 Installation and operation of rotor 0 this part is optional 1 Rotor 0 consists of a fixing sleeve rotor 0 and a testing tube...

Page 6: ...ture is controlled and the instrument is leveled 6 Rotor 0 cannot run with no load when no liquid is filled and the rotor protective frame is unnecessary when rotor 0 is used Fig 5 Fig 6 Fig 7 8 Steps...

Page 7: ...cording to the sample viscosity range Notice the universal joint must be gently lifted up when attaching or removing the extension hook and the rotating shaft must be protected from damage Fig 8 Fig 9...

Page 8: ...bar and fix the other end of the rotor sleeve to the central axis Transfer a suitable amount of the tested sample into the material tube See details in Fig 11 2 Rotate the elevating knob to lower the...

Page 9: ...select and hook a rotor according to the sample viscosity range See details in Fig 15 5 According to the selected rotor take a corresponding amount of sample tested see details in Tab 1 into the mater...

Page 10: ...ce this key is pressed UP The corresponding set value will be increased by 1 once this key is pressed 2 The vertical bars under the set value represent the heating power percentage of machine output a...

Page 11: ...eparation before test 1 For series NDJ SNB LVDV viscometers transfer the tested sample into a round and flat bottom container whose diameter is 60mm at least For series RVDV HADV HBDV viscometers the...

Page 12: ...rument will firstly display the initial interface and then enter the main menu after 3sec and the main menu has 4 option bars Viscosity test test the viscosity of the sample View results view and prin...

Page 13: ...he previous testing conditions before power off Quick test test the viscosity rapidly on by selecting a rotor and a rotating speed without saving the tested data New test create a new test with a new...

Page 14: ...gh Continue test in the next test This group of parameters can be selected from the Stored test in future testing Fig 20 1 4 Start testing after parameters are well set click Run to start testing and...

Page 15: ...the left top to return to the list interface and click Print to print the current test data and coordinate curve through an externally equipped printer an external printer is required Fig 22 1 6 Clic...

Page 16: ...ow in which the red dot lines at both ends represent the minimum value and the maximum value E g the standard viscosity of the sample is 12 500cp 17 500cp in the figure below the actually measured vis...

Page 17: ...e Scan and when the scanning result satisfies the test requirement Suitable will be displayed and then click Use to apply the scanning result in the testing parameters If the scanning result cannot sa...

Page 18: ...meters please read as per the model or configuration of the instrument you purchased 18 Fig 27 4 Setting set the time and date etc of the instrument Fig 28 Fig 28 4 1 Time and date setting click Time...

Page 19: ...light adjustment click Backlight to adjust the screen brightness via the and or by directly dragging the circular adjusting icon Fig 30 Fig 30 4 3 Reset to the factory parameters Click Reset to enter...

Page 20: ...tion bars Viscosity test test the viscosity of the sample View test data view and print the saved test result Manage test data output or delete the saved test data in batch Setting set the basic param...

Page 21: ...isplayed as a curve in a coordinate system After testing click Stop to complete the test or testing will be automatically terminated at the end of the testing time set If the sample viscosity is exces...

Page 22: ...ertain linear area so the torque percentage shall be controlled during measurement and this value shall be 10 90 and if the angle percentage is too high or too low ERROR will be displayed for the torq...

Page 23: ...t is determined by the liquid property and is not any problem arising from the instrument VIII Viscometer standard packing list SN Name Qty SN Name Qty 1 Digital display viscometer main machine 1 set...