Atmos E 1 Operating Instructions Manual Download Page 1

MedizinTechnik

ATMOS

®

 Chair E 1

GA1GB.130203.0
2019-06  Index: 07

Operating Instructions

English

Summary of Contents for E 1

Page 1: ...MedizinTechnik ATMOS Chair E 1 GA1GB 130203 0 2019 06 Index 07 Operating Instructions English ...

Page 2: ... up 7 3 3 Electrical connection 8 4 0 Operation 9 4 1 Positioning the patient 9 4 2 Adjusting the seat height 9 4 3 Rotating the upper part of the chair 10 4 4 Adjusting the backrest 10 4 5 Adjusting the arm rests 10 4 6 Adjusting the head rest 11 4 7 Foot rest 11 4 8 Chassis optional 11 5 0 Cleaning and care 12 5 1 General information on cleaning and disinfection 12 5 2 Cleaning and disinfection ...

Page 3: ...r to start up please peruse chapter 2 0 For your safety in order to be prepared for any possible dangerous situations 1 3 Function The electrically height adjustable ATMOS Chair E 1 patient chair has been designed for medical examinations Height adjustment is achieved by an encapsulated infinitely variable electric actuator The removable headrest is also height adjustable The moving and fold away ...

Page 4: ...ions Please press where dot indicates Subnumeration Numeration General information Follow the arrows Replace Check Please read important information Move plug in this direction Engage check correct fit Turn shift in this direction Important information Symbols of ATMOS Chair E 1 Order number This product complies with the relevant requirements of EU Directive Degree of protection type B REF Replac...

Page 5: ...mperature for at least six hours If the device is not acclimatized it may not be used as damages to the electronic components may be the result Only transport the device in a shipping carton which is padded and offers sufficient protection If damage occurs during transport Document and report the transport damage Send the device to ATMOS Chapter 6 2 Sending in the device on page 14 Ambient conditi...

Page 6: ...ersonal injury and damage to property if no original ATMOS parts are being used the advice for use in these operating instructions is not being observed The correct installation and handling of the device is essential for safe operation Therefore prior to starting up you must become thoroughly familiar with the contents of these operating instructions The electric motor is protected by an integrat...

Page 7: ...e surface 3 2 Starting up Position the chair at the allocated space Any floor unevenness must be compensated for Check that the upper part of the chair can rotate freely Peruse safety information in part 2 0 prior to starting up the device for the first time Finally connect power cable ...

Page 8: ... outlet features a proper ground connection The electrical connection values voltage and nominal frequency as well as the data for fuses can be found on the type plate above the connecting socket To disconnect the device pull the mains plug out of outlet socket There is no indication that the device is powered Disconnect chair if not in use during service and repair work and or for cleaning 3 0 Se...

Page 9: ...ection switch After 1 5 minutes of continued operation the motor requires a cool down period of approx 8 5 minutes If the thermal protection switch is activated the motor requires a cool down period of approx 20 minutes 4 1 Positioning the patient Ensure that your patient sits in the middle of the seat A constant unilateral strain on the seat can damage the surface 4 2 Adjusting the seat height Th...

Page 10: ...osition Release lever element which will then return to its initial position Backrest and foot support are arrested Arm rests base part and back rest are coupled for synchronous movements Fig 2 Lever element 4 3 Rotating the upper part of the chair Upper part rotates around 360 via a ball bearing and is equipped with an arrestor each 90 The upper part with the patient can then be rotated in the de...

Page 11: ... the green LED illuminates fig 6 Push the lever to the left fig 8 to extend the chassis Move the chair downwards until the red LED illuminates fig 6 The chair can now be moved Retract the chassis Move the chair upwards until the green LED illuminates fig 6 Push the lever forward fig 7 to retract the chassis Move the chair to the desired position LEDs fig 6 Red Extend the chassis chair can be moved...

Page 12: ...self can be wiped off with a moist not wet cloth Substances such as blood need to be removed immediately to prevent stains Do not use aggressive or abrasive cleansing agents For the upholstery usual dry foam may be used Treatment with a commonly used care product for artificial leather is recommended once a week in order to keep the upholstery soft and smooth The described action relating to clean...

Page 13: ... following active ingredients QAV quarternary ammonium compounds 5 4 Recommended disinfectants for the upholstery Disinfectant Ingredients in 100 g Manufacturer ATMOS Green Clean SK Application solution Alkyl dimethyl benzyl ammonium chloride 1 g Metasys Rum Austria Dialkyl dimehtyl ammonium chloride 1 g Alkyl dimethyl ethyl benzyl ammonium chloride 1 g ...

Page 14: ...ion Possible cause Remedy Device does not start Power plug is fitted badly Check connection at wall socket No mains voltage Check main fuse Check mains plug on perfect fit at the device Defect fuse Exchange of fuse Should a malfunction still occur please inform the authorized ATMOS customer service immediately 8 0 Accessories and spare parts REF 8 1 Accessories Child seat On request 8 2 Spare part...

Page 15: ...e current Max 0 1 Ω Max 0 5 mA Max 0 1 mA Max 0 1 mA Ambient conditions transport storage Temperature 10 50 C Humidity without condensation 30 95 Air pressure 500 1060 hPa Ambient conditions for operation Temperature 10 35 C Humidity without condensation 30 95 Air pressure 500 1060 hPa Dimensions H x W x D 1420 x 630 x 920 mm Weight 92 kg Weight with movable base plate 100 kg Period tests Repeat t...

Page 16: ...e device may not be disposed of in domestic waste Please observe existing national laws and rules for disposal of old devices in the respective country Disposal within the Federal Republic of Germany In the Federal Republic of Germany the law for electrical devices ElektroG regulates the disposal of electrical devices In order to guarantee a proper disposal of your old device please either pass on...

Page 17: ...tact 8 kV Air Floors should be made of wood or concrete or tiled with ceramic tiles If floors are synthetic the relative humidity should be at least 30 Fast electrical transient burst IEC 61000 4 4 2 kV Mains 1 kV I Os 2 kV Mains 1 kV I Os Mains power quality should be that of a typical commercial or hospital environment Surges IEC 61000 4 5 1 kV Common 2 kV Differential 1 kV Common 2 kV Different...

Page 18: ...1 3 Guidelines and Manufacturer s Declaration Immunity The ATMOS Chair E 1 is intended for use in the electromagnetic environment specified below The customer or user of the ATMOS Chair E 1 should ensure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Conducted disturbances acc to IEC 61000 4 6 3 Veff 150 kHz to 80 MHz...

Page 19: ...f 150 kHz to 80 MHz the field strength should be below 3 V m The ATMOS Chair E 1 is intended for use in electromagnetic environment in which radiated disturbances are controlled The customer or user of the ATMOS Chair E 1 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications equipment and the ATMOS Chair E 1 as recommended bel...

Page 20: ...ATMOS MedizinTechnik GmbH Co KG Ludwig Kegel Straße 16 79853 Lenzkirch Germany Phone 49 7653 689 370 atmos atmosmed de www atmosmed com MedizinTechnik ...