Atlas Weather-Ometer Ci3000+ Service Manual Download Page 11

 / Ci4000 / Ci5000 Weather-Ometer

®

 Service Guide   Doc. No. 07-7030-00  Rev. 0

34

5.0 TROUBLESHOOTING PROCEDURES

5.2 Xenon Lamp and Irradiance Control Malfunctions (cont.)

5.2.2 Troubleshooting Guide

MALFUNCTION

PROBABLE CAUSE

CORRECTIVE ACTION

Lamp fails to ignite

Chamber door open or switch failure

Close door or check switch.

Xenon lamp/ignition breakers off

Turn on circuit-breakers.

Lamp assembled/installed incorrectly

Check lamp.

Lamp ignitor lead off or loose

Check lead connections.

Light rod dirty or damaged

Inspect/clean light rod.

Light monitor malfunction

Check filters & photodiodes.

Lamp burner socket damaged

Inspect/replace (Sec. 7.2.4).

Low lamp starting voltage

Check/adjust taps on T1 and L1.

CS-6 cooling water flow low

Check CS-6 (see Section 5.5)

Ignitor malfunction

Conduct procedure in 5.2.6.

Relay failure in lamp circuit

Conduct procedure in 5.2.6.

PLC malfunction

Conduct procedure in 5.2.6.

Lamp output is high/low

Burner is near end of life

Replace xenon burner.

Lamp filters expired

Replace lamp filters.

Burner tip/socket damaged

Inspect/replace socket or burner.

Incorrect irradiance calibration

Recalibrate (Section 6.9).

Interference filters expired

Check/replace filters (Sec. 7.2.9).

Light rod is misaligned

Align light rod (Secs. 7.2.14 & .15)

Wattage regulator malfunction

Conduct procedure in 5.2.7.

Lamp will not operate

1-Sun or 2-Sun relay failure

Perform procedure 5.2.8.

above 1-Sun level
(Ci5000 only)

Lamp operating at

Burner is near end of life

Replace xenon burner.

excessive wattage

Lamp filters expired

Replace lamp filters.

Burner tip/socket damaged

Inspect/replace socket or burner.

Incorrect irradiance calibration

Recalibrate (Section 6.9).

Interference filters expired

Check/replace filters (Sec. 7.2.9).

Light rod is misaligned

Align per Section 7.2.14 or 7.2.15.

Wattage regulator malfunction

Conduct procedure in 5.2.7.

No wattage display

Blown fuse F1 on A1 transducer

Check fuse.

A1 wattage transducer board failure

Check board output w/ lamp on.

Summary of Contents for Weather-Ometer Ci3000+

Page 1: ...Ci3000 Ci4000 Ci5000 Weather Ometer Instrument Service Guide Weathering Instrumentation Products Doc No 07 7030 00 Revision 0 November 10 1999 Weather Ometer Ci3000 Ci4000 Ci5000...

Page 2: ...1 4 Troubleshooting Procedure 25 5 1 5 Power Component Configuration 26 5 2 Xenon Lamp and Irradiance Malfunctions 29 5 2 1 Description 29 5 2 2 Troubleshooting Guide 34 5 2 3 Troubleshooting Flow Ch...

Page 3: ...3 Humidification Nozzle rear panel 48 Figure 5 3 4 Humidification Nozzle Detail 48 Figure 5 3 5 Humidifier Water Line Filter 48 Figure 5 3 6 Humidification Block Diagram 49 Figure 5 3 7 Humidificatio...

Page 4: ...e 8 5 8 AFE Data Panel Mark I early Ci4000 193 Figure 8 5 9 AFE Data Panel Mark II 193 Figure 8 5 10 Typical PLC Module side view 194 Figure 8 5 11 PLC0 Fuse 195 Figure 8 6 1 Ci4000 Xenon Lamp Ignitor...

Page 5: ...Sensor BPT and BST Chamber dry bulb Sensor Relative Humidity Lamp Wattage Lamp Irradiance Air valve Damper Rack Speed Control Resistivity Monitor Lamp Rack Alignment n Maintenance Lamp Water Fitting...

Page 6: ...e main reactor contribute additional power and provide precision wattage control Irradiance control is provided by the light monitor which continuously monitors lamp output and provides a voltage sign...

Page 7: ...RELAY K9 4 3 F1 FUSE INTERNAL 062A 250V TP1 TP2 A1 2 A1 1 A1 5 APC1 32 31 APC1 34 33 PLC5 8 A1 4 PLC5 10 A1b WATTAGE TRANSDUCER 1 9 10 WATTAGE TRANS DUCER RELAY K9 RESISTOR 500 OHM PLC5 ANALOG INPUT M...

Page 8: ...TION CIRCUIT BREAKER CB2 3 4 IGNITION PULSE RELAY K3 6 5 2 1 BLEEDER RESISTOR 56K OHM 3 4 3 4 C2 C3 BOOST CAP RELAY K4 56 56 XENON CIRCUIT BREAKER CB1 1 2 3 4 BOOSTER TRANSFORMER T1 2 3 1 XENON CIRCUI...

Page 9: ...2 IGNITION CIRCUIT BREAKER CB2 IGNITION PULSE RELAY K3 IGNITOR L3 HIGH VOLTAGE TRANSFORMER T4 6 7 5 3 4 2 1 6 w b brn w b blk 3 4 9 10 3 2 7 BLEEDER RESISTOR 56K OHM IGNITION CAPACITOR C1 80 MFD WATT...

Page 10: ...r a 1 1 mixture of alcohol and acetone Hold the lamp by the upper or lower filter housing only Check that there is a sealing washer inside the lamp water fitting Connect the lamp and rotate it until t...

Page 11: ...ct procedure in 5 2 6 Lamp output is high low Burner is near end of life Replace xenon burner Lamp filters expired Replace lamp filters Burner tip socket damaged Inspect replace socket or burner Incor...

Page 12: ...gnition pulse is audible Install calibrated lamp and attempt ignition Lamp ignites Lamp does not ignite Previous lamp is malfunc tioning Inspect for dam age Check burner socket Measure line voltage an...

Page 13: ...n circuited or wattage regulator failed No continuity Check for 24Vac across relay K9 coil terminals Voltage present No voltage present Replace relay K9 If B5 LED on PLC3 is on replace PLC3 If LED is...

Page 14: ...d relay K15 terminal K15 3 If voltage present K15 or K14 is stuck closed Replace bad relay 4 Check for 24Vac at K14 coil terminals K14 1 and K14 2 Voltage present No voltage Replace relay K14 5 With K...

Page 15: ...High is displayed reset the chamber safety thermostat S4 inside the electronics drawer Press the red reset button on the thermostat 4 Initiate standard test no 1 to attempt lamp ignition 5 If the igni...

Page 16: ...nce Control Malfunctions cont 5 2 6 Troubleshooting Procedure Lamp Ignition Failure Ignitor cont Ci3 4000 Terminals K3 5 and K3 6 Ci5000 Terminals K3 6 and K3 7 A If there is continuity across the K3...

Page 17: ...ator cont terminals PLC5 11 and PLC5 13 or at terminals PLC5 12 and PLC5 14 if the optional sensor is installed and is controlling irradiance A If no voltage is present there is a problem with the lig...

Page 18: ...ked in a later step 9 Stop the wattage test and alter it to the mid range operating wattage level shown below Ci3000 3000 watts Ci4000 5000 watts Ci5000 8000 watts 10 Initiate the new test to ignite t...

Page 19: ...7 Total Segments 1 Duration Units Time Black Sensor BPT Black Temp Active Yes Radiation filter 340 nm Chamber Temp Active No Segment Data Panel page 8 Segment Number 1 Light or Dark Light Time or Irr...

Page 20: ...sed Replace the malfunctioning relay Sec 8 1 B If there is no continuity check for 24Vac at the K14 coil terminals K14 1 and K14 2 If voltage is present replace K14 If no voltage is present and the A8...

Page 21: ...arrow band interference filter is different than the one used with a 300 400nm wide band filter Do not mix them Photodiode Part Numbers 340nm 420 nm 300 400 nm Older Ci4000 09 3515 00 09 3515 00 09 34...

Page 22: ...2 8 Chamber Overtemperature Thermostat This thermostat is accessed from inside the electronics drawer but may be easier to replace by removing the right side access panel Guidelines 1 The face of the...

Page 23: ...ram 69 Malfunction Troubleshooting 69 Motor Controller Input Jumper 75 Blower Wheel Mounting 73 BP Transmitters Collector Rings 78 Burner Socket Maintenance 158 C Calibration 134 Black Sensor Check 13...

Page 24: ...l Diagrams 103 Malfunction Troubleshooting 99 Plumbing Ci3 100 Plumbing Ci4 Ci5 99 Water Solenoid Valves 100 Wattage Setting Minimum 40 Wattage Calibration 148 Wattage Powermeter Connections calibrati...

Reviews: