Atlas SUNTEST XLS Service Manual Download Page 13

 

 

Service manual SUNTEST XLS/ XLS+ 

Page 13 

4.  Service software 
 

In the main display you can call the Service software with key „9“.Then you are asked for a keyword. 
Please enter „1901557“. 
Now you can use the following functions with the corresponding number keys: 
 

„1“: Pt 1000- Test 

 
The display shows the analogue inputs of the two Pt- 1000 sensors for BST and CHT: 
Display: 

xxx (5-255bit) A/D 

 

xxx BST- temperatur (0-185°C) 

   

 

xxx (5-255bit) A/D 

 

xxx CHT- temperatur (0-185°C) 

 

 
„2“: Digital Input- Test 
 

The display shows the four different digital inputs. 
 
Display: 

x(0/1) door switch 

 

 

 

 

x(0/1) status lamp on/off 

   

 

x(0/1) Max. power on power supply board x(0/1) thermo switch immersion                  

                                                                                              unit 
 
„0“- display means not active. „1“- value means, that the input is active (+24VDC). 
 

 
„3“: Digital Output- Test 
 

The display shows the four digital outputs. With „Enter“ and „Escape“ the different functions can be 
switched on and off. 
 
Display: 

x(0/1)  lamp on/off 

 

 

x(0/1) immersion on/off 

   

 

x(0/1)  No function 

 

 

x(0/1) lamp cooling on/ off

 

 
 
„4“: PWM1- Test (PWM= pulse width modulation) 

 
Here you can check the function of the control for chamber fan. 
 
Display: 

xxx (0-256bit) Chamber fan min/ max speed 

 
With „arrow up“ and „arrow down“ you can change the PWM- value.  
 

 
„5“: PWM0- Test 

 
With this function you can check the PWM- control for the lamp. 
 
Display: 

xxx (160-800bit) lamp min/max power 

 
With „arrow up“ and „arrow down“ you can change the PWM- value for the lamp. 

Summary of Contents for SUNTEST XLS

Page 1: ...ram XLS XLS 3 Components 3 1 Power Supply Board Lamp 3 2 Ignition Unit 3 3 Control Board 4 Service Software 5 Adjustment Calibration 5 1 Adjustment on power supply board lamp with Xenocal Radialux 5 2...

Page 2: ...s must be protected against the Xenon light via protecting glasses f e sun glasses Please take care of yourself and the environment in case of using chemical or biological cleaning and greasing substa...

Page 3: ...w5 w25 T2 Blatt 1 3 PE3 N PE PE4 F F E D Vorschaltger t mit Lichtsensor und Triac Z ndger t mit Entst rdrossel X 1 1 X 2 1 X 2 2 X 3 1 X 3 3 X 1 4 A2M 5 w3 3 s w w3 4 s w X 1 3 K1 M 470 Triac G A 1 A...

Page 4: ...4 h B St1 1 St1 2 St1 3 1 S2F 1 4 X 4 1 w18 br X 7 1 w18 bl w16 bl w16 br X 6 7 X 7 2 1 4 X 1 X 5 K1M 1 3 M2M PE N L sw b l gnge A1M B Frontplatte vst Hauptschalter 3 L1 A w14 br 1 Deckel schalter 1L...

Page 5: ...Service manual SUNTEST XLS XLS Page 5 3 Components 3 1 Power Supply Board Lamp Circuit diagram power supply board lamp 1 2...

Page 6: ...Service manual SUNTEST XLS XLS Page 6 Circuit diagram power supply board 2 2...

Page 7: ...Service manual SUNTEST XLS XLS Page 7 Layout components side power supply board...

Page 8: ...Service manual SUNTEST XLS XLS Page 8 3 2 Ignition Unit Ignition unit...

Page 9: ...Service manual SUNTEST XLS XLS Page 9 Circuit diagram ignition unit...

Page 10: ...Service manual SUNTEST XLS XLS Page 10 3 3 Control board Circuit diagram control board 1 2...

Page 11: ...Service manual SUNTEST XLS XLS Page 11 Circuit diagram control board 2 2...

Page 12: ...Service manual SUNTEST XLS XLS Page 12 Layout components side control board...

Page 13: ...0 1 Max power on power supply board x 0 1 thermo switch immersion unit 0 display means not active 1 value means that the input is active 24VDC 3 Digital Output Test The display shows the four digital...

Page 14: ...ta will be erased This function resets the whole memory to default values all customer s programs will be erased clock 0 00 operating hours 0 arrow up filter factors The display shows the saved intern...

Page 15: ...pot TP4 fully clockwise Select the chamber fan speed to maximum at 256 PWM by using the arrow keys Turn pot TP4 fully clockwise Select program 5 Select program 5 Start Lamp Start lamp Start Lamp Start...

Page 16: ...to the instruction manual can be done f e when customers have different filter systems No software adjustment possible If necessary a software calibration according to the instruction manual can be d...

Page 17: ...Service manual SUNTEST XLS XLS Page 17 5 2 Adjustment E via Software...

Page 18: ...Service manual SUNTEST XLS XLS Page 18...

Page 19: ...Service manual SUNTEST XLS XLS Page 19 5 3 Adjustment BST via Software...

Page 20: ...Service manual SUNTEST XLS XLS Page 20 6 Application Standards 6 1 Tables of specifications...

Page 21: ...Service manual SUNTEST XLS XLS Page 21...

Page 22: ...Service manual SUNTEST XLS XLS Page 22 6 2 Diagrams...

Page 23: ...Service manual SUNTEST XLS XLS Page 23...

Page 24: ...Service manual SUNTEST XLS XLS Page 24 7 Immersion unit See operating manual of the unit itsef...

Page 25: ...witched on off via control panel For the Suntest CPS or XLS an additional PCB must be connected between the unit and the suncool The suncool unit is not controlled It cannot be used to control any tem...

Reviews: