- 19 -
Operating Manual SunSpray for SUNTEST
®
XLS+
6
Start up
6.8 Incoming power connection
WARNING – electric shock!
Touching live parts can lead to life threaten-
ing electric shocks. Check plugs and power
cables for damage before connecting to the
incoming power. Damaged parts may not be
used for electrical connection!
The SunSpray device is connected to a incoming voltage
of 230 V ± 10 % (see Technical Data). The incoming power
connection is made by a PE plug (1P/N/PE) or (2P/PE).
Device fuse 10 A (F). The electrical power must be fused
according to regulations.
Connecting and turning on the device:
1. Before connecting to the incoming power, check
whether the values of the power supply network of
the operating rooms match the data on the rating plate
on the back of the device.
2. If the data for voltage (V) and maximum current (A)
do not match, the device may not be connected.
3.
Fig. 6.3:
Insert the PE plug
1
(on the back of the
device) with one side into the light equipment socket
2
and the other side into a properly grounded and
fused socket.
4. Make sure that no stress or strain forces act on the
power cables.
5. The spraying cycle of the SunSpray is controlled by
the program control of the SUNTEST
®
XLS+. Plug and
fix the connecting cable
3
(9-pin D-Sub plug) into the
connection on the back of the SUNTEST XLS+ for this.
6. Turn on the SunSpray device at the switch.
Fig. 6.3
1
2
3