Appendix D - Troubleshooting
Problem
Possible Solution
Capacitance measured when
probes are open circuit is not
close to zero (±1.0pF).
Perform a probe compensation.
Resistance and/or inductance
measured when probes are
short circuit is not close to
zero (±1.2Ω, ±1.6µH).
Perform a probe compensation.
Measured value doesn’t
appear to be correct.
Ensure probes are well connected to the
component under test for the entire
duration of the analysis.
Ensure that nothing else is connected with
the component under test. Make sure that
you are not touching the connections.
The component value may be outside the
supported measurement range.
The component’s design frequency may
not correspond to the test frequencies used
by the
LCR40
.
Measured values vary
slightly between tests.
The displayed resolution is higher than the
measurement resolution to avoid rounding
errors. Small variations within the quoted
measurement resolutions are normal.
Calibration date is
approaching or is in the past.
Don’t worry, the
LCR40
will carry on
working after the “Calibration Due Date”.
The date is simply a recommendation.