Export/import calibration
Export
<cr>
<cr>
<cr>
export calibration string from calibrated device
Import
import calibration string to new device
Export,?
calibration string info
Command syntax
<cr>
<cr>
<cr>
<cr>
<cr>
59 6F 75 20 61 72
(1 of 10)
(1 of 10)
(2 of 10)
(8 more)
(10 of 10)
(FIFO)
65 20 61 20 63 6F
6F 6C 20 67 75 79
*DONE
<cr>
Import, 59 6F 75 20 61 72
<cr>
Export
<cr>
Export
<cr>
Export
<cr>
Export
10,120
Disabling *OK simplifies this process
Export,?
Example
Response
Export: Use this command to save calibration settings
Import: Use this command to load calibration settings
to one or more devices.
Import, n
Response breakdown
# of strings to export
Export strings can be up to 12 characters long,
and is always followed by
<cr>
10,
# of bytes to export
120
r 0.1
27