Summary of Contents for BYK 5415

Page 1: ...Measure what you see byko cup M Manual...

Page 2: ......

Page 3: ...H Lausitzer Str 8 D 82538 Geretsried Germany Tel 0 800 gardner 0 800 4273637 49 8171 3493 0 Fax 49 8171 3493 140 BYK Gardner USA 9104 Guilford Road Columbia MD 21046 USA Phone 800 343 7721 301 483 650...

Page 4: ...4 English Contents Table of contents 1 Information 5 2 System Description 7 3 Start up 8 4 Operation 10 5 Maintenance 14 6 Components Ordering Guide 16 7 Technical Data 17...

Page 5: ...sm expose to excessive heat or humidity aggressive and or solvent and or corrosive substances ammable substances excessive vibration Defects and Extreme Usage If safe operation can no longer be presum...

Page 6: ...byko cup M Zero plate Clamp ring Operating handle Button LCD display Base plate Magnifying glass Information Clamp ring Operating handle Base plate LCD display Zero plate Button Operating handle Magni...

Page 7: ...ear box for ease of use Double handed operation Integral digital readout Zero retention at power off Ergonomic design for ease of use and viewing of test 10x illuminated magni er 20mm range with 0 01m...

Page 8: ...ation with adequate access to the front and sides 2 The power adapter is supplied with plugs for various country requirements Select the plug suitable for your requirements Insert the power adapter in...

Page 9: ...lamp knob Adjust to a position for optimum viewing of the sample 7 Carefully install the two handles into the clamping ring Turn the handles on the clamping ring counterclockwise to raise the clamping...

Page 10: ...ero the instrument s measurement system zero position The display will show Touch Zero Plate Then Press Button Lightly hold one of the operating handles with two ngers and slowly turn the handle count...

Page 11: ...s Choose one of the two procedures below Method A Pass Fail A pre determined depth of deformation is applied to the test panel in order to test for compliance to a speci cation The test panel may or m...

Page 12: ...rument and remove these before starting another test 10 Turn the handles in a counterclockwise direction as viewed from the right hand side of the instrument to lower the hemispherical indenter until...

Page 13: ...s range of 0 3mm to 1 25mm maximum Do not exceed the above gures as this may cause permanent damage to the instrument Zero All measurements are taken from the zero position where the indenter is level...

Page 14: ...ing handles for complete travel of the indenter nominally 20 mm Ensure the movement is smooth and even 3 Inspect the hemispherical indenter for any signs of wear for example ats on the surface Replace...

Page 15: ...ve the instrument serviced and calibrated by a trained and quali ed service engineer Cleaning Do not use solvent to clean the outer surfaces of the instrument Use a damp cloth and mild detergent solut...

Page 16: ...JIS B 7729 Cat No Description 5415 byko cup M 5416 Zero Plate 5417 Magni er 5418 Indenter Service and Spare Parts For all service and spare parts requirements please contact either BYK Gardner USA 910...

Page 17: ...inches 280 mm Spherical Punch 0 8 inches 20 mm Accuracy 0 002 inches Full Range 0 05 mm Full Range Full Travel 0 02 0 81 inches 0 5 20 5 mm Gearing 1 revolution of handle moves punch 0 2 mm under loa...