BYK-Gardner GmbH
Acquire Color Compact
300 400 159 - 2107
7
2 System Operation
2.1 Instrument Check
A color check on
green
is recommended every
7 days
, calibration on
white
is
forced every
30 days
.
Check instrument on color tile
1.
Select
Check
in the main menu.
2.
Place unit on
green
color tile and tip on
OK
or press button
Operate
.
Values are within tolerance
1.
Confirm success message with
OK
.
2.
The unit returns to the main menu and readings can be taken.
Values are out of tolerance
1.
Confirm error message with
OK
and repeat
green
color tile reading.
2.
If the color check is not successful for three times, you are asked to per-
form a calibration on
white
tile and thereafter, a color check on
green
tile
again.
If calibration fails, clean the
white
tile. If calibration still fails, contact customer
support.
2.2 Sample Measurement
In this mode free measurements can be made and named. Display of measure-
ment values after finishing can be activated through the PC. Select
Measure
in
the main menu and place instrument on sample to be measured.
Summary of Contents for 345 000 035
Page 47: ...BYK Gardner GmbH Acquire Color Compact 300 400 159 2107 47 2 2 1 7 30 1 2 OK 1 OK 2 1 OK 2 2 2...
Page 48: ...Acquire Color Compact BYK Gardner GmbH 48 300 400 159 2107 1 2 3 4 5 2 3 Sample 1 2 OK 3 X OK...
Page 49: ...BYK Gardner GmbH Acquire Color Compact 300 400 159 2107 49 2 4 1 2 OK 3 4 5 6 2 5 1 2 3 OK...
Page 50: ...Acquire Color Compact BYK Gardner GmbH 50 300 400 159 2107 2 6 1 2 3 4 OK 5 2 7 1 Std 2 3 4 OK...
Page 57: ...BYK Gardner GmbH Acquire Color Compact 300 400 159 2107 57 2 4 1 2 OK 3 4 5 6 2 5 1 2 3 OK...
Page 58: ...Acquire Color Compact BYK Gardner GmbH 58 300 400 159 2107 2 6 1 2 3 4 OK 5 2 7 1 Std 2 3 4 OK...