Part 2 – Analyzer Mounting
2.1
General
All Q46 Series instruments offer maximum mounting flexibility. A bracket is included with each
unit that allows mounting to walls or pipes. Choose a location that is readily accessible for
calibrations and keep in mind that it will be necessary to use solutions during the calibration
process. To take full advantage of the high contrast display, mount the instrument in a location
where the display can be viewed from various angles and long distances.
Locate the instrument in close proximity to the point of sensor installation - this will allow easy
access during calibration. The sensor-to-instrument distance should not exceed 100 feet (30 m).
To maximize signal-to-noise ratio however, work with the shortest sensor cable possible. The
standard cable length of the turbidity sensor is 30 feet.
Figure 1
– Q46 Enclosure Dimensions