Q46H/65 Residual Chlorine Dioxide System
Part 1 - Introduction
7
O&M Manual
Rev-J (11/18)
Figure 3 below shows an installation using a 00-1522 sealed flowcell for the ClO
2
sensor and a 00-1527
sealed flowcell for the pH sensor. This type of installation requires careful flow control. We recommend
the use of our 03-0372 flow control assembly when using sealed flowcells. This assembly consists of an
in-line filter and a fixed-flow regulator which will maintain a constant 400 cc/min flow rate through the
system. This flow will be maintained so long as inlet pressures are between 5 and 125 PSIG. The in-line
filter is used mainly to protect the flow control element against larger particles that might cause plugging
of the device.
Figure 3 - Sealed Flowcell Assemblies with Flow Control