8
O&M Manual
Rev-L (7/15)
Part 3 –Electrical Connection
3.1
General
The sensor cable can be quickly connected to the Q45terminal strip by matching the wire
colors on the cableconductors. A junction box is also available to provide a quick-
disconnect point for the sensor, or a break point for very long sensor cable runs. Route
signal cable away from AC power lines, adjustable frequency drives, motors, or other
noisy electrical signal lines. Do not run sensor or signal cables in conduit that contains
AC power lines or motor leads.
3.2
DirectSensorConnection
The sensor cable should be routed into the enclosure through the right hand cord grip on
the bottom of the enclosure. Make sure the cord-grips are snugly tightened after
electrical connections have been made to prevent moisture incursion. When stripping
cables, leave adequate length for connections in the transmitter enclosure, as shown
below.
Figure 2 - Sensor Cable Preparation