Trig-Tek™ 

530B 

Tracking Filter 

User Manual 

 

Publication No. 980974  Rev. B 

 
 
 
 
 
 
 
 
 
 
 
 

Astronics Test Systems Inc. 

4 Goodyear, Irvine, CA 92618 

Tel:  (800) 722-2528, (949) 859-8999; Fax: (949) 859-7139 

 

 

[email protected]    [email protected] 

[email protected]    http://www.astronicstestsystems.com 

 

 

 

Copyright 2011 by Astronics Test Systems Inc.  Printed in the United States of America.  All rights reserved.  
This book or parts thereof may not be reproduced in any form without written permission of the publisher. 

Summary of Contents for Trig-Tek 203M

Page 1: ...9 859 8999 Fax 949 859 7139 atsinfo astronics com atssales astronics com atshelpdesk astronics com http www astronicstestsystems com Copyright 2011 by Astronics Test Systems Inc Printed in the United...

Page 2: ...tion s of any product or part without Astronics Test Systems express written consent or misuse of any product or part The warranty also does not apply to fuses software non rechargeable batteries dama...

Page 3: ...y procurement of products from Astronics Test Systems TRADEMARKS AND SERVICE MARKS All trademarks and service marks used in this document are the property of their respective owners Racal Instruments...

Page 4: ...ion cord or a three prong two prong adapter This will defeat the protective feature of the third conductor in the power cord Maintenance and calibration procedures sometimes call for operation of the...

Page 5: ...Level Log Optional 1 5 DC of Frequency Log 1 5 Bandwidth 1 5 Time Constant 1 6 Controls 1 6 Indicators 1 6 Power 1 7 Size 1 7 Cabinets required 1 7 Chapter 2 2 1 Operation 2 1 Meter Input Switch 2 1...

Page 6: ...i Astronics Test Systems Chapter 4 4 1 Calibration Procedure 4 1 Test Equipment 4 1 Test Procedure 4 1 DC of Frequency Adjustments Linear 4 2 DC Zero Adjustments W15B 4 3 FS Range Adjust 4 3 Log of Ad...

Page 7: ...o 980974 Rev B 530B User Manual Astronics Test Systems iii List of Figures Figure 1 1 530B Tracking Filter 1 2 Figure 4 1 Adjustment Resistor Locations 4 2 Figure 4 2 Front Panel Meter Circuit Board A...

Page 8: ...530B User Manual Publication No 980974 Rev B iv Astronics Test Systems This page was left intentionally blank...

Page 9: ...Rev B 530B User Manual Astronics Test Systems v DOCUMENT CHANGE HISTORY Revision Date Description of Change A 03 31 2011 Document Control release B 01 28 2012 Revised per ECN02268 updating bandwidth...

Page 10: ...530B User Manual Publication No 980974 Rev B vi Astronics Test Systems This page was left intentionally blank...

Page 11: ...Hz the filter BW will stay at 10 of the reference frequency until it reaches 500 Hz where the bandwidth would be 50 Hz Thereafter as the reference frequency increases the filter bandwidth would remain...

Page 12: ...neral The 530B Tracking Filter has two major parts The Carrier Generator provides tuning to the analyzer portion and generates a linear and logarithmic DC voltage proportional to frequency The Analyze...

Page 13: ...e 0 and 90 MDAC in the W15 where the signal is mixed with the carrier signals and translated to a DC IF Intermediate Frequency The DC plus side band signal is filtered by the variable Low Pass filter...

Page 14: ...round Filtered Signal Output Impedance 50 Ohms Frequency 5 Hz to 10 000 Hz Level 0 to 10 Volts RMS Distortion 5 to 10 000 Hz 5 20 to 5 000 Hz 2 Accuracy 3 of Reading 0 5 FS Dynamic Range 50dB Connecto...

Page 15: ...ctor BNC Low side at signal ground DC of Level Log Optional Impedance 50 Ohms Maximum Output 10 Volts 0dB Sensitivity 125 mV dB Dynamic Range 50dB WBW NBW Optional Noise 50dB Connector BNC Low side at...

Page 16: ...o Limit mode Bandwidth is 10 of the REF frequency In the MANUAL mode a fixed bandwidth is selected via the front panel BW Hz thumb switch Filter IN OUT Switch Located on the front panel of the chassis...

Page 17: ...30 Volts 50 400 Hz Size 7 High x 2 7 Wide x 13 Deep Cabinets required Single Module 4220 Six Module 19 Rack 4200 Optional Plug in HP and LP Filter Assy 1697 1 is available with 3 selectable High Pass...

Page 18: ...530B User Manual Publication No 980974 Rev B Introduction 1 8 Astronics Test Systems This page was left intentionally blank...

Page 19: ...01 to 299 in 1 Hz steps The WB input has a HP LP Filter Assembly option with settings available to filter the WB output See WIDE BANDWIDTH FILTER at the end of this section The PH DEG selects the DC o...

Page 20: ...t an overload condition exists and the acquired data may not be valid ERROR Light This indicator is a red LED labeled ERROR The ERROR LED illuminates when an incorrect signal is at the REF Input Corre...

Page 21: ...O switch to MANUAL the BW HZ thumb switch to 005 and the DC OUTPUT TC switch to SLOW Connect the power cord of the rack assembly into the power line and turn the POWER switch at the left side of the r...

Page 22: ...V on the DC voltmeter 21 Observe that the front panel meter indicates 10 00 0 25 22 Connect the AC Voltmeter to the FILTER OUTPUT jack 23 Observe a 10 00 0 5 VRMS indication on the AC voltmeter 24 Se...

Page 23: ...et the frequency of generator 2 to 500 0 1 Hz and the level to 1 000 0 005 VRMS 6 The front panel meter should indicate 1 00 03 VRMS 7 Set generator 2 for 501 Hz 8 Observe 0 707 0 1 VRMS indication on...

Page 24: ...530B User Manual Publication No 980974 Rev B Performance Test 3 4 Astronics Test Systems This page was left intentionally blank...

Page 25: ...hed W15 board Even though they may have the same resistor number be sure to adjust the correct resistor Test Equipment Note Equivalent equipment can be substituted Multimeter Fluke 4588A Generators 2...

Page 26: ...S3 to LIN c S4 to LIN d S5 to OFF e S6 to SE Figure 4 1 Adjustment Resistor Locations DC of Frequency Adjustments Linear 1 Connect the generator to the REF INPUT jack Pins 17 and 18 of P1 2 Set the ge...

Page 27: ...point S2 7 Set R2 DC ADJ for a 0 3 mV indication on the DC voltmeter FS Range Adjust 1 Connect the generator to both of the REF and DATA INPUT jacks Pins 17 and 18 and Pins 13 and 16 of P1 2 Place th...

Page 28: ...the RANGE switch to 10V and the generator for 10 00 05 VRMS 4 Set R3 NB LOG ADJ for a 10 00 0 1 V indication on the DC voltmeter 5 Connect the DC voltmeter the WB OUTPUT jack Pin 14 of P1 6 Set R4 WB...

Page 29: ...ER INPUT switch to NB the BANDWIDTH switch to MANUAL and set the BW HZ thumb switch to 200 5 Set the frequency of generator 2 to 1000 0 1 Hz and the level to for a 1 000 01 volt indication on the LEVE...

Page 30: ...530B User Manual Publication No 980974 Rev B Calibration Procedure 4 6 Astronics Test Systems This page was left intentionally blank...