background image

AC Current

 

Range 

Resolution 

Accuracy 

400µA 

0.1µA 

± (1.5% of rdg + 8dgts) 

4000µA 

1µA 

40mµA 

10µA 

400mµA 

100µA 

4A 

1mA 

 

± (2.0% of rdg +10dgts) 

10A 

10mA 

Overload Protection:   µA and mA ranges: F0.5A/600V fuse 
 

 

 

4A and 10A ranges: F10A/600V fuse 

Max. Input Current: 

“INPUT” Terminal: 200mA 
“A” Terminal: 10A 
(For measurements >5A: only measure in durations less than 10 seconds, 
in intervals greater than 15 minutes.) 

Voltage Drop:  400µA, 40mA and 4A ranges: 40mV 
 

 

4000µA, 400mA and 10A ranges: 400mV 

Frequency Range: 40Hz ~ 400Hz 
Response: Average, calibrated in rms of sine wave 
 

Resistance (Auto Ranging)

 

Range 

Resolution 

Accuracy 

400Ω 

0.1Ω 

± (1.5% of rdg + 3dgts) 

4KΩ 

1Ω 

40KΩ 

10Ω 

400KΩ 

100Ω 

4MΩ 

1KΩ 

40MΩ 

10KΩ 

Open Circuit Voltage: ~0.25V 
Overload Protection: 250V DC/AC rms 

Summary of Contents for DT132A

Page 1: ...if all or part of the exterior casing is removed Look for cracks or missing plastic Pay special attention to the insulation around the connectors Inspect the test leads for damaged insulation or expos...

Page 2: ...and damage Turn the Meter off when not in use and take out the battery when it is not going to be used for an extended period of time Regularly check the battery as it may leak when it has not been u...

Page 3: ...al Polarity Indication displayed Overload Indication OL displayed Low Battery Indication Yes Operating Environment 32 F 104 F 0 C 40 C 80 RH Storage Temperature 14 F 122 F 10 C 50 C 85 RH Power 3 x 1...

Page 4: ...de when measuring voltage current resistance or capacitance When the meter is in auto range mode AUTO is displayed on the screen To enter or exit manual range mode press the RANGE button When entering...

Page 5: ...maximum and minimum values will be cleared after cleared when the test function changes or the meter is turned off 6 Relative Test Press this button to store the current reading and resets the display...

Page 6: ...t Impedance 10M Overload Protection 600V DC or 600V AC rms 200mV range 250V DC AC rms Max Input Voltage 600V DC AC Voltage Auto Ranging Range Resolution Accuracy 400mV 1mV 1 2 of rdg 5 dgts 4V 1mV 1 0...

Page 7: ...ed for standard configuration DC Current Range Resolution Accuracy 400 A 0 1 A 1 2 of rdg 8dgts 4000 A 1 A 40m A 10 A 400m A 100 A 4A 1mA 10A 10mA Overload Protection A and mA ranges F0 5A 600V fuse 4...

Page 8: ...l 200mA A Terminal 10A For measurements 5A only measure in durations less than 10 seconds in intervals greater than 15 minutes Voltage Drop 400 A 40mA and 4A ranges 40mV 4000 A 400mA and 10A ranges 40...

Page 9: ...it voltage about 0 5V Overload Protection 250V DC AC rms For continuity test When the resistance is between 50 and 100 the buzzer may sound or may not sound When the resistance is more than 100 the bu...

Page 10: ...d 10A connect the red test lead to the 10A jack instead 2 Set the range switch to desired A mA or A range If the current magnitude to be measured is not known beforehand set the ranges switch to the h...

Page 11: ...ead to the anode of the diode to be tested and the black test lead to the cathode 5 The meter will show the approximate forward voltage of the diode If the connections are reversed OL will be shown on...

Page 12: ...he specified ratings Then reinstall the battery cover and lock the cover DISPOSAL OF THIS ARTICLE If you at some point intend to dispose of this article please keep in mind that many of its components...