ASTRO-PHYSICS 1100GTO Manual Download Page 31

30

POLAR ALIGNMENT OPTIONS – FINE POLAR ALIGNMENT

 

For very casual visual observation, you may skip this section and simply start observing.  A finder-scope or single power 

finder may be required to locate objects since GoTo slews with the keypad require good polar alignment for spot-on 

accuracy.  Don’t forget to tighten your Altitude Lock Knobs (4) before you begin!  For casual visual observation, you can 

move the telescope manually via the clutches or by using the GoTo and centering functions of the keypad.   

The keypad and GTO Servo Control Box will function as soon as they are plugged in.  That means that the R.A. axis will be 

moving westward at the sidereal rate.  Note, however, that this motion will not be accurate enough to qualify as “tracking” 

without accurate polar alignment.  In short, if you will depend on the GoTo functions of the 1100GTO, or if you intend to do 

astrophotography, you must perform a more accurate polar alignment.  Some methods, procedures and tips are presented 

below.  You will complete this alignment when your scope and other equipment are mounted. 

Methods for fine polar alignment

 

Polar Alignment Scope 

– Our optional Right-Angle Polar Alignment Scope (RAPAS) will allow you to quickly align 

your mount on the pole.  The reticles are designed for use in both the Northern and Southern hemispheres.  Even 

users of the GTO computerized mounts will find these polar scopes useful, particularly if your telescope is not 

orthogonal to the mount (please refer to the keypad manual for a discussion of orthogonality).  If you have a polar 

alignment scope, please read the instructions sheets that come with it.  If you are planning long-exposure astro-

imaging, we suggest that you use the polar alignment scope, then tweak the final polar alignment by using traditional 

drift alignment, the GTO Quick Star-Drift Method (refer to the Keypad v4.19.3 manual), 

PEMPro

™ Pole Align Wizard 

or other similar alignment program. 

 

GTO Keypad

 – Please refer to the instruction manual for the GTO Keypad and read the sections from “Getting 

Started” through “Alternate Polar Calibration Routines & Tips.”  As time goes on, the keypad manuals will be updated.  

Please refer to the Technical Support section of the website for the most recent documentation.  Here are summary 

descriptions of several techniques for polar alignment from the current Keypad Manual available on the Technical 

Support page of our website.  

 

The Keypad startup routine provides two methods:  The North Polar Calibrate and the Two Star Calibrate.   

These two polar alignment methods, though no longer recommended, were designed for quick and coarse 

alignment in the field with portable setups.  They are for visual observers, not imagers.  The Two Star Method 

is the better of the two as it is less affected by the extremely low resolution in R.A. near the pole and by 

orthogonality issues.

 

The Daytime Routine (see “Polar Aligning in the Daytime” in the Keypad Manual), is a great trick for daytime 

setup.  It will allow you to “wow” your friends by setting up and finding planets and bright stars in the daytime. 

In addition, it is the recommended first step in alignment for anyone in the southern hemisphere.  Even those 

in the south who own our polar scope will find it helpful, since it will generally put the rather difficult-to-spot 

southern stars into the polar scope’s field of view.  

 

The GTO Quick Star Drift Method is for use with a finder scope.  By using a finder scope, you are able to 

remove orthogonality issues from the process, making subsequent alignments much easier. 

 

 

Using the Daytime Routine, followed by the GTO Quick Star Drift Method, will provide accurate enough 

polar alignment for extensive imaging.  This combination is our recommended procedure for anyone in the 

southern hemisphere, or anyone who finds their view of the pole obstructed.

 

Computer Software Solutions

 

 

Polar Align via PEMPro’s “Pole Align Wizard” – PEMPro

 is a software by Ray Gralak and distributed by 

CCDware.  This powerful application not only analyzes and improves the periodic error performance of any 

mount that is equipped with a CCD camera and compatible camera control software, but also includes “Pole 

Align Wizard”.   

 

“Pole Align Wizard” is one of the best method of accurately polar aligning!   It is like traditional drift aligning, 

but on steroids!  It has the advantage of automatically choosing the optimized stars for drift aligning and does 

not have to worry about star magnitude since the CCD can see dimmer than an eye.  Also, it compensates for 

air refraction and other variables that manual drift aligning cannot.  It does require the use of a CCD camera 

and corresponding camera control software.  The Wizard walks you through the alignment process with easy 

step-by-step instructions.  

Note:  This and the R.A. Correction Method (on the Technical Support page 

of our website) are the preferred alignment method when tweaking the RAPAS orthogonality. 

 

A full PEMPro

 edition is included with the 1100GTO, 1600GTO and 3600GTO mounts (sorry, not the 

Mach1GTO

). 

Summary of Contents for 1100GTO

Page 1: ...in Starting in March 2021 March 2021 Keypad Optional Astro Physics Inc Astro Physics Inc 11250 Forest Hills Road 11250 Forest Hills Road Machesney Park IL 61115 Machesney Park IL 61115 Telephone 815...

Page 2: ...VING SITE 13 Assemble Pier purchased separately 13 Attach Polar Axis Assembly to Pier 13 Latitude Adjustment for 1100GTO German Equatorial Mounts 14 Special 90 Degree Alt Azimuth Position 15 ASSEMBLY...

Page 3: ...3 Lanyard Post and Bubble Level 33 Warning for Dec Servo Cable 33 SLEWING YOUR MOUNT IN BELOW FREEZING TEMPERATURES 33 AUTO ADJUSTING GEARBOX NO MORE WORM MESHING 34 Checking and Adjusting the Gearbox...

Page 4: ...ation and for future updated versions of this manual A final note and an apology to our friends in the southern hemisphere Many of the instructions in this manual are written entirely from the point o...

Page 5: ...Cable tie wraps 1 USB Flash Drive containing APCC Standard PEMPro v3 Software and PDF of Instruction manuals Instruction Manuals and Registration Card Fasteners 4 1 4 20 x 5 8 SHCS for safety backup o...

Page 6: ...accessories camera computer dew removers etc You may wish to consider CP4 Powerpoles to Alligator Clips CABPPAL For connecting the CP4 Power Cable to battery terminals DC In line Watt Meter and Power...

Page 7: ...weight 12 9 lb 5 8 kg Optional shaft extension available Weight of mount R A axis polar fork 27 6 lb 12 5 kg Dec axis 15 6 lb 7 1 kg Dec top plate 1 5 lb 0 7 kg Capacity of mount Approximately 110 lb...

Page 8: ...xtensive individual testing that is performed on each and every mount While the native peak to peak periodic error of your 1100GTO will be 7 arcseconds or less you can reduce it even further to maximi...

Page 9: ...o this mount It is important to note the need to connect and disconnect the Dec motor servo cable when doing so Control Box Adapter located behind CP4 GTOCP4 Control Box Removal Thumb Screws 2 Declina...

Page 10: ...onnect the Dec motor servo cable when doing so Declination Motor Gearbox Dec Sighthole Cover Declination Axis Housing Declination Counterweight Adapter Counterweight Shaft Safety Stop Altitude Adjuste...

Page 11: ...eypad instructions in the Keypad manual These four points are detailed below 1 Latitude and longitude coordinates tell the mount where it is The GTOCP4 will remember where it was when last pow ered do...

Page 12: ...e 10 After finishing your night s observing or day s solar observing it is best to park the mount into your chosen Park posi tion by pressing 2 Setup then 4 Park Mount Opt and finally choosing the Par...

Page 13: ...Start Guide 6 Power up the mount and connect with your primary control software Use the primary control software to initialize the mount Start out from the park position chosen in 5 above Your other i...

Page 14: ...r pier or into the 1100 Flat Surface Adapter 119FSA if are using one 4 Line up the Pier base If the side thru holes in the pier or 119FSA are not perfectly lined up with the tapped holes in the mount...

Page 15: ...uding from both sides of the bar This will allow you to move the mount fully within the altitude range The slots are positioned at 9 degree increments and the threaded rod has 10 degrees of travel The...

Page 16: ...reverse the axis and the other to remove and replace the bolts 1 Remove telescopes and counterweights before beginning this orientation change It is also necessary to separate the Dec axis so that yo...

Page 17: ...ns the two servo cable branches will be attached to the GTOCP4 box at the connector marked motors 1100GTO AE AEL The 1100GTO AE and 1100GTO AEL Absolute Encoder versions of the mount utilize a more co...

Page 18: ...der mount is pre routed through the R A axis The Dec Servo Cable Y cable of non AE mount will need to be routed through the R A axis by the user Note that there is also a re movable Cable Router provi...

Page 19: ...late to the axis and lock it in place by hand tightening the two lock knobs You may wish to place and tighten the two 1 4 20 x 5 8 socket head screws that provide additional attachment security It is...

Page 20: ...le they do not actually need to be set up on a telescope This will also allow you to have an idea about cable length if there are choices to be made Once you have all your devices connected and know t...

Page 21: ...scope 4 Begin by feeding the cable with the largest connector from the front of the axis to the rear You will need to finesse the cable through the rear Cable Access opening If you are hav ing troubl...

Page 22: ...by first feeding the cable with the largest connector through the Cable Access opening at the bottom rear of the R A axis You will need to finesse the cable through the opening and up through the cent...

Page 23: ...1 4 20 x 5 8 socket head cap screws 18 FLAT MOUNTING PLATE FP1800 This plate is 18 long and 7 5 at its widest point in the center The width of the plate tapers to 5 5 at each end Four pairs of keyhole...

Page 24: ...so that the saddle can be shifted forward or backward depending on whether you are balancing a front heavy or back heavy scope In addition to D style plates it matches well with the Astro Physics 15...

Page 25: ...ars In order to retain the tilt in feature of the dovetail the sliding bars must have an approximate width at the widest point between 1 65 42 mm and 1 8 45 mm and they must have have a 75 degree beve...

Page 26: ...lutch Knobs section before attaching your telescope This will give you an idea of how the clutch knobs feel and function Flat and ribbed plates Our flat and ribbed plates are constructed with keyhole...

Page 27: ...ackward and forward With a well balanced telescope the above tightness of the clutch knobs will be sufficient for all normal conditions of use Now if you proceed to mount up and balance your telescope...

Page 28: ...ing Precision Balancing Again begin in Park 3 Remember that dangling cables will dramatically change balance and create guiding problems so you ll want to be sure that all cables are carefully secured...

Page 29: ...r weights to the top of the shaft and try balancing with just the lower lighter weight If you must slide the weights lower as necessary to balance however it is better to add another heavy weight up h...

Page 30: ...azimuth adjustment knobs one on each side of the mount to center it You must back off the opposing azimuth knob in order to move the other knob in that direction One full turn of the azimuth knob is a...

Page 31: ...bsite The Keypad startup routine provides two methods The North Polar Calibrate and the Two Star Calibrate These two polar alignment methods though no longer recommended were designed for quick and co...

Page 32: ...ecause our GTO Quick Star Drift Method begins with altitude Many texts for the classic star drift method begin with the azimuth adjustments When you made your rough alignment earlier you loosened ever...

Page 33: ...k off the first knob only by the small amount of the adjustment you plan to make Use the graduated markings on the knob to mark your starting and ending points For example if you are already pretty cl...

Page 34: ...n some observatory situations use a heavy wire to minimize the voltage drop If the power drops below about 10 5 11 volts at the servo terminal the internal computer chips may reset with subsequent los...

Page 35: ...d scope in line with the R A axis pointing toward the pole 2 Remove the Lever Assembly Cover Grip the cover and pull it straight out 3 Make sure the lever is in its normal position snapped in place pa...

Page 36: ...he entire motor gearbox is securely fastened to the axis housing There should be a very small amount of front to back rocking that is possible due to the spring loading mechanism but there should be n...

Page 37: ...ory or at a week long star party then unpark from Last Parked assuming that the scope has not been repositioned manually by hand via the clutches since it was last parked Unpark from a defined park po...

Page 38: ...ff marks from repeated transport and assembly disassembly Most of the time these marks can be removed with a product like Color Back by Turtlewax automotive product Simply apply with a paper towel and...

Page 39: ...ded for best performance Do NOT exceed a nominal 18 volt system See the Power Considerations section beginning on page 29 If you also power other equipment CCD cameras dew heaters etc from the same so...

Page 40: ...parts of the sky Ray Gralak s Pulse Guide will allow you to dial in an exact tracking rate for any part of the sky The Astro Physics Command Center Pro APCC includes tracking and pointing correction...

Page 41: ...Y axis and he is absolutely certain that Y Dec After much sending of equipment back and forth to our facility it finally is determined that it was the other axis after all that had the problem PLEASE...

Page 42: ...l know how to fix it Once you know that the mount is tracking in a normal fashion with normal periodic error profile you can go from there to begin setting up your guiding parameters It is not a piece...

Page 43: ...enter North pause South pause North to center The stars in the resulting exposure should look something like the image to the right The scale might be different but you should see what looks like many...

Page 44: ...t in the reverse direction It is normal on reversal of direction that there is a slight delay in movement because of backlash If this happens you may wish to increase the pulse count If you see errati...

Page 45: ...late 2 1 4 20 x 1 2 SHCS for center hole in rings 1 1 4 20 x 3 8 SHCS for safety stop SBV15 15 Wide Profile V Style Dovetail Plate 2 1 4 20 x 1 2 SHCS for center hole in rings 4 1 4 20 x 3 4 FHSCS for...

Reviews: