Parameter setup and measurement
LDS70A User Manual
Page 30
ASTECH GmbH
▪
Distances are always entered in meter (m) with 3 decimal places after
the separator.
631 cm
Input: 6.310
5.123 m Input: 5.123
▪
Retrieve parameter values
Input:
PARAMETER<CR (0x0D)>
▪
Set parameter values
Input:
PARAMETER VALUE1 VALUE2<CR (0x0D)>
▪
Starting a measurement (operating mode)
Input:
COMMAND<CR (0x0D)>
▪
Stopping a measurement (continuous operating mode)
Input:
<ESC (0x1B)>
The set parameters are stored in the non-volatile EEPROM immediately. The
last entered data will be available upon restarting.
5.3
Identification
ID - Device identification information
When entering the command ID, the LDS70A will respond by displaying the
device identification information in the order ID, serial number, firmware
version.
Query:
ID<CR (0x0D)>
Response:
ID SN 180004 V3.38R 630 (example)