page 41 of 48
7.2
ASSKEA pro
wound
®
S
Flow rate
max. 8 l/min
Pressure
max. -200 mmHg;
Conversion factor: 1 kPa ~ 7.5 mmHg
Canister
Disposable exudate canister system (1000 ml)
Suction tube
Disposable suction tube, Ø 4 mm (internal),
length 180 cm, sterile (REF: 100712)
Power supply unit
FRIWO FW 7555M/12, cable length 4 m
Nominal voltage of the power supply
unit
In: AC 100 – 240 V~ / 50-60 Hz / 350 – 150 mA
In (UL only): 120 Vac / max. 350 mA / 50-60 Hz
Out: DC 12 V / 1,25 A
Maximum load current
1.25 A
Nominal voltage of the circuit board 12 V
Power consumption at 12 V
15 W
Current consumption at 12 V
1.25 A
Protection class as per IEC 60601-1
Type BF
Risk classification as per 93/42/EEC,
IX
IIa
Protection class as per IEC 60601-1
II
Degree of protection (IP code) as per
IEC 60529
IP22
CE marking
CE1434
UL marking
Sound emission
Operation:
35 dB (A)
High priority alarm:
53 dB (A)
Low priority alarm:
51 dB (A)
Ambient conditions
Transport/Storage:
-25°C to +60°C
humidity of max. 93%
non-condensing
Operation:
+5°C to +40°C
humidity 15% to 93%
non-condensing
Air Pressure:
700 hPa to 1060 hPa
Battery, rechargeable
7.4 V, 4.4 Ah – Lithium ions
Charging time if battery is empty
Charging time if battery is approx.
50% full
6 - 7 hours
3 - 3.5 hours
Dimensions (H x W x D) in mm
290 x 259 + 100 (canister) x 130
Weight (base unit)
2.2 kg
Pressure measurement accuracy
Target pressure > -80 mmHg
max.
5%
Target pressure < -80 mmHg
max.
10%
Operating time
Continuous operation
Battery-powered runtime
approx. 24 – 48 hours, depending on the run time of the
motor
Item number
100513-3
AS TO ELECTRICAL
SHOCK, FIRE AND
MECHANICAL HAZARDS
ONLY IN ACCORDANCE
WITH ANSI/AAMI
ES60601-1 (2005),
CAN/CSA-C22.2 No.
60601-1 (2008)