40
ENGLISH
Housing
Phenomenon
Test method
Immunity test levels
Electrostatic discharge
IEC 61000-4-2
± 8 kV contact
±15 kV air
Radiated RF disturbances
IEC 61000-4-3
10 V/m
80 MHz to 2,7 GHz
80 % AM at 1 kHz
Radiated RF disturbances
pursuant to frequencies
and immunity test levels of
EN 60601-1-2, table 9
see EN 60601-1-2, Table 9
Magnetic field at rated
power frequency (50 Hz)
IEC 61000-4-8
30 A/m
Power supply AC
Phenomenon
Test method
Immunity test levels
Fast transient electrical
disturbances/bursts
IEC 61000-4-4
±2 kV
100 kHz repetition frequency
Impulse voltages /surges
IEC 61000-4-5
±1 kV line-to-line
±2 kV line-to-ground
Conducted RF disturbances
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V in ISM bands and amateur
radio bands between 0.15 MHz
and 80 MHz
80 % AM at 1 kHz
Voltage dips
IEC 61000-4-11
0 % U
T
for ½ period
At 0°, 45°, 90°, 135°, 180°, 225°,
270°, 315°
0 % U
T
for 1 period
70 % U
T
for 25 periods
one-phase at 0°
Voltage interruptions
IEC 61000-4-11
0 % U
T
for 250 cycles
Note: U
T
is the AC mains voltage prior to the application of the test levels.
8.1 Electromagnetic environment, in
which the ASSKEA secretion as-
pirators M20 and S-series may be
operated
The ASSKEA secretion aspirators M20
and S-series are intended for operation in
the electromagnetic environment speci-
fied below, in which RF disturbances are
controlled. The customer or the user of the
ASSKEA secretion aspirators M20 and S-
series must ensure that they are operated in
such an environment.
The environments for intended operation
include professional healthcare institutions
and the home care environment. Special en-
vironments, such as close to RF surgery or
MRI or environments, in which the intensity
of EMC disturbances is high, are excluded.
Emission limits
Conducted and radia-
ted RF emissions
CISPR 11