Please note that, the pump can not be
started beyond an existing vacuum of
-0.5 bar when switching the device on
again.
Always take note of additional im-
portant instructions in section 2.3.3
regarding handling of the ASSKEA
double filter system!
Ensure that the battery has a sufficient
capacity during the aspiration proce-
dure. The battery status display must
signal green or yellow. If the display
turns red, the external power supply
unit must be plugged in immediately.
In this case, aspiration can continue
without interruption.
3.5 Replacement of the disposable
liner ‘OneWay‘
Health risks associated with handling
infectious agents or pathogens
Infectious agents and pathogens in
the aspirate cause health risks.
• Wear suitable disposable gloves
when replacing the disposable liner
‘OneWay‘.
• Use the disposable liner ‘OneWay‘
for one patient only.
• Replace the disposable liner ‘One-
Way‘ and suction tube in accordance
with the respectively applicable hy-
giene instructions, no less than once
per week, prior to each new patient
or if it is full.
• For each new patient, profession-
al hygienic processing by ASSKEA
GmbH or an authorized service part-
ner of ASSKEA GmbH is strictly re-
quired!
• After each aspiration procedure, all
components that have come into
contact with the aspirate must be
cleaned, disinfected or disposed of.
• Disposal of aspirate and contaminat-
ed components must be performed
in a proper manner.
1. Switch off the ASSKEA secretion aspira-
tors M20 resp. S-series by setting the pow-
er switch (fig. 1(F), fig. 2(F)) to ‘O‘.
2. Separate the suction tube from the patient
connection.
3. Separate the pre-assembled connecting
tube of the disposable liner ‘OneWay‘ at
the bottom end of the T-piece of the exter-
nal canister ‘Bag‘.
4. Remove the disposable liner ‘OneWay‘
from the reusable external canister ‘Bag‘.
5. Dispose of the disposable liner ‘OneWay‘
and suction tube in a proper manner (see
section 6.3).
6. Insert a new disposable liner ‘OneWay‘
in the reusable external canister ‘Bag‘ as
specified in section 3.2.2. Ensure that the
connecting tube as well as the lid of the
disposable liner are properly seated on the
external canister.
7. Attach a new suction tube to the patient
connection of the disposable liner ‘One-
Way‘.
8. Switch on the ASSKEA secretion aspirator
M20 resp. S-series.
If the reusable secretion canister sys-
tem is used, please follow the instruc-
tions regarding cleaning /disinfection
in section 4.1.3 or 4.2.3 and 4.1.4 or
4.2.4.