16
2
1
NOTE:
The below table indicate the suggested CLP pressure according to the patients weight.
This is intended as a guide only as patient body shapes vary.
Constant Low Pressure
CLP mode offers an immersion low pressure surface operating between 15mmHg to 29mmHg depending
on comfort settings chosen.
1. Press the
MODE
button until the
CLP
LED light is illuminated
CPL PRESSURE CHART
SWL
250KG
COMFORT CONTROL
INDICATORS
Mode/Output Pressure
(mmHG)
PATIENT WEIGHT RANGE (KG)
Alternate
Static
CLP
30
60
80
100
125
160
200
250
15
15
15
20
20
17
25
25
19
30
30
21
35
35
23
40
40
25
45
45
27
50
50
29