25. ELECTROMAGNETIC COMPATIBILITY
Medical electrical equipment is subject to special precautions with regard to EMC and must be installed and
commissioned in accordance with the EMC instructions contained in the accompanying documents.
No special measures are to be observed for devices and systems of ASP GmbH.
Portable and mobile HF communications equipment can affect medical electrical equipment.
Guidelines and manufacturer's declaration
– Electromagnetic immunity (Table 201)
The product is intended for operation in the ELECTROMAGNETIC ENVIRONMENT specified below. The
customer or the user of the product should ensure that it is used in such an environment.
Transmission measurement
Agreement
Electromagnetic Environment - Guidelines
High-frequency (RF) transmissions
according to CISPR 11
Group 1
The product uses RF radiation exclusively
for internal functions. For this reason, the RF
radiation of the device is very low, and it is
unlikely that neighboring electrical devices
will be disturbed.
High-frequency (RF) transmissions
according to CISPR 11
Class B
The product is intended for use in all
facilities, including residential areas and
those directly connected to a public utility
network that also supplies buildings used for
residential purposes.
Harmonics according to IEC 61000-3-2
Class A
Voltage fluctuations / Flicker after
IEC 61000-3-3
Jibes
Summary of Contents for motioncare LIANA
Page 15: ...Gebrauchsanweisung LIANA Version 1 0 Seite 15 von 50 7 1 Product dimensions...
Page 39: ...Gebrauchsanweisung LIANA Version 1 0 Seite 39 von 50 21 ELECTRICAL DIAGRAM...
Page 40: ...Gebrauchsanweisung LIANA Version 1 0 Seite 40 von 50...
Page 42: ...Gebrauchsanweisung LIANA Version 1 0 Seite 42 von 50 23 CE DECLARATION OF CONFORMITY PSS LIANA...
Page 50: ...Gebrauchsanweisung LIANA Version 1 0 Seite 50 von 50...