Installation Guide
teleCARE IP
Peripherals
7.8
Room Display (NIRD)
The Room Display (NIRD) can be connected to the NICB for the use of two sensor inputs.
An 8-pole screw terminal is used for the connection of the Room bus and the NICB.
Figure 165. Room Display
Note:
The 8-pole connector terminal which is required to connect the room bus
and the external bus to the NICB module must be ordered separately. The article
number is R180732.
7.8.1 Connecting to the NICB
The NICB cross cable needs to be prepared to be connected to the 8-pole screw terminal
before connecting to the Room Display module.
Note:
The cross cable wires are not solid and cannot be used with the NICT8-AA
8-pole connector.
• Cut off the end of the cross cable.