Operation and Frequently Used Functions
Synergy
RF™
User’s Guide
DFU-0221-6r0_fmt_en-US
Page 39 of 58
Figure 8
Information Screen - Operation
Table 14
Information Screen Display - Operation
1.
Model, Software Versions & Builds
for each controller.
2.
Menu
button
Returns to the menu screen.
3.
Home
button
Returns to the main screen.
5.2.2.1 Arthrex Apollo
RF®
device (Probe) Limits - IDR
The Arthrex
Synergy
RF
generator has built in intelligent device recognition (IDR)
for Arthrex Apollo
RF®
probes. When attached, the recommended setting will be
automatically applied, but individuals that desire an alternate setting can use the
ABLATE
power,
COAG
power, or
Ablate power change button
(see Figure 13)
on the device to change the settings. Some power settings may be not be
allowed, depending on the RF probe attached
5.2.3
Reset Defaults Screen
The
Reset Defaults
screen prompts the user to confirm resetting to factory
defaults. This screen is displayed when the Reset Defaults button is pressed on
the
Settings
screen.
3
1
2
Summary of Contents for Synergy AR-9800
Page 2: ...This page intentionally left blank ...
Page 26: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 24 of 58 ...
Page 27: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 25 of 58 ...
Page 28: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 26 of 58 ...
Page 29: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 27 of 58 ...
Page 30: ...Technical Specifications SynergyRF User s Guide DFU 0221 6r0_fmt_en US Page 28 of 58 ...