DrillSaw Mini 300™ System
User’s Guide
DFU-0223-2 Rev. 0
Page 11
3
3
.
.
0
0
Technical Specifications
3.1
Handpiece
Table 4
Handpiece AR-300
Width
94 mm (3.7”)
Height
47 mm (1.85”)
Depth
175 mm (6.889”)
Weight
550 g (1.21 lbs.)
Cannulation
2.4 mm (0.0945”)
Protection
IPX0
Speed
See Section 3.2 Table 7 Attachments
Torque
See Section 3.2 Table 7 Attachments
Applied part type
B
Pollution level
2
Overvoltage category II
Altitude
Up to 3,000 m [9,842 ft.] above sea level
Duty Cycle*
Operating Type S3 per IEC 60034-1
At 25°C (77°F) Ambient Temperature:
30 sec on / 30 sec off for 5 cycles, then wait 48 minutes
*
Following the duty cycle is an important part of maintaining the Arthrex Mini 300
Power System. The duty cycle times have been determined under average load,
using the Sagittal Saw attachment and at ambient air temperature of 25° C (77° F).
Operating the device above of these prescribed parameters can lead to user and
patient injury, as well as a shortened life span of the device itself. The recommended
environmental operating temperature is 25° C (77° F) or lower. Operating in
environments above 25° C (77° F) can shorten the above prescribed duty cycles. The
environment temperature, surgical attachment type and time need to be considered
by the user, when planning its use in surgical interventions. Additionally, please
read the
General Warnings and Safety Notices
that start on Page 1 of this User’s Guide.
Table 5
Ambient conditions for operation
Temperature
10° to 40 °C (50° to 104 °F)
Relative Humidity
15% to 80% (relative), non-condensing
Table 6
Ambient conditions for storage (in shipping packaging)
Temperature
-40° to +70°C (-40° to +158°F)
Relative Humidity
8% to 80% (relative), non-condensing