Cleaning and Disinfecting
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 30 of 43
Cleaning and Disinfecting
6.0
6.1
Console (AR-6480)
The AR-6480 console is provided non-sterile and should not be sterilized.
The AR-6480 console can be cleaned/disinfected using a cloth and commercially
available surfactants or surface disinfectants only. The AR-6480 must not be
submersed in any liquid.
Always comply with the instructions issued by the manufacturer of
the disinfectant.
NEVER use liquid to clean the remote control connector contacts.
Remove dust regularly with dry compressed air.
6.2
Remote Control Unit (AR-6482)
The Remote Control Unit (AR-6482) is supplied
non-sterile
.
The Remote Control Unit can be autoclaved for sterilization.
See Section 7.0 for sterilization information.
The Remote Control Unit can be cleaned/disinfected using commercially
available surfactants or surface disinfectants. Always comply with the
instructions issued by the manufacturer of the surfactant or disinfectant. It is not
designed to be submersed in Gluteraldehyde, Steris®, or Sterrad® disinfectants.
NEVER use liquid to clean the connector contacts of the remote
control unit connector. Remove dust regularly with dry
compressed air.
6.3
Footswitches Control Unit (AR-6483)
Clean the footswitch with an enzymatic cleaner without subsequent acid
neutralization.
Rinse the footswitch thoroughly after cleaning.
After cleaning, disinfect the footswitch with a commercially available surface
disinfectant.
Thoroughly rinse the footswitch under lukewarm water.
Always comply with the instructions issued by the manufacturer of
the disinfectant.
NEVER allow the console receptacles to have any contact with
liquids. If there is dust or moisture on the receptacles, remove with
dry compressed air. ONLY dry connectors should be plugged into
the console.