950-0073-00 B SM, English, Arthrex Synergy UHD4 System
20 of 74
2.0
System Installation
and Operation with
Data Input Device
2.1
Installation
NOTE
:
1. Your Synergy
UHD4
Camera Control Unit will
indicate which software configuration is
enabled at boot up
on the Video Monitor’s
Splash screen.
2.1.1
Typical System Installation
NOTE: See Typical Interconnect Diagram,
Figure
NOTE: Synergy
UHD4
may be installed in a
tower or on an equipment boom.
1. Place Synergy
UHD4
Camera Control Unit
(CCU) on tower shelf or installed on
equipment boom.
2. Attach monitor to the tower or equipment
boom and connect monitor DC power cable
to the rear panel of the monitor as shown.
3. Attach Synergy
UHD4
Data Input Device to
secondary tower arm or equipment boom.
Connect the cable from the Data Input
Device
to the connector labeled “tablet” on
the back of the Synergy
UHD4
CCU.
4. Connect a Display Port cable to a Display
Port output on the rear panel of the
Synergy
UHD4
CCU. Connect the other end
of the Display Port cable to the Display Port
input of the display monitor. (3G-SDI or DVI
cables may be used instead of Display Port
cables.) Note: Arthrex recommends
connecting Synergy
UHD4
CCU to the primary
surgical monitor via multiple output types
(e.g., Display Port and DVI, Display Port and
3G-SDI) in the event that one type of
connection is lost.
5. If using a printer, connect the printer cable to
the USB connector on the rear panel of the
Synergy
UHD4
CCU. Connect the other end
of printer cable to the printer.
6. Plug the AC power cord into the
Synergy
UHD4
power inlet module and a
standard grounded AC Mains outlet (100-
240 V˜, 50-60Hz).
7. Insert the card edge connector of the
Synergy
UHD4
camera head into the camera
receptacle on the front of the CCU.
NOTE: Ensure the camera head connector
contacts are clean and dry prior to
insertion.
8. Connect the Light Guide cable into the Light
Guide receptacle on the front panel of the
Synergy
UHD4
CCU. Attach the other end of
the Light Guide cable to the endoscope.
10. Insert the endoscope into the Synergy
UHD4
camera head grasping mechanism or into
the C-Mount Adapter for C-Mount Heads.
11. Press the Light Source On/Standby Switch
to activate the LED light engine.
NOTE
: If there is no Light Guide cable
connected to the Synergy
UHD4
CCU,
pressing the On/Standby Switch will not
activate the LED light engine until one is
connected.